Entris
®
II Advanced Line Operating Instructions
21
Operating Concept
4 .11 .4 Parameters in the “Settings / Weighing” Menu
Parameters
Settings
Explanation
Safety Level
High*
If the balance is
not
correctly leveled or a calibration/adjustment is not needed
or the minimum initial weight requirements according to USP are
not
met:
−
A warning message is displayed .
−
No
data is transmitted .
−
The printout is blocked .
−
Starting and saving in applications is blocked .
Standard
If the balance is
not
correctly leveled or a calibration/adjustment is not needed
or the minimum initial weight requirements according to USP are
not
met:
−
A warning message is displayed .
Low
−
If the balance is
not
leveled: The [Start leveling] button is active .
−
If the balance needs to be calibrated and adjusted: The [isoCAL] button is
active .
Ambient
conditions
Very stable
Sets the ambient conditions to “very stable” . Activates a fast change in the
weight values in the event of a load change with a high output rate .
Recommended for the following work environment:
−
Very stable table near the wall
−
Closed and calm room
Stable*
Sets the ambient conditions to “stable” . Recommended for the following work
environment:
−
Stable table
−
Slight movement in the room
−
Slight draft
Unstable
Sets the ambient conditions to “unstable”: Activates the delayed change in
weight values with a reduced output rate . Recommended for the following work
environment:
−
Simple office desk
−
Room with moving machinery or personnel
−
Slight air movement
Very unstable
Sets the ambient conditions to “very unstable”: Activates a significantly delayed
change in the weight values and long wait for stability with a further reduction in
the output rate .
Recommended for the following work environment:
−
Noticeable and slow floor vibrations
−
Noticeable building vibrations
−
Weighed goods moved
−
Very strong air movements
Application
Weighing*
Activates a filter that enables a fast change in the display for very fast load
changes . Display changes with minimal load changes (in the digit range) occur
more slowly .
Dosing
Activates a filter that enables a very fast change in the display with minimal load
changes, e .g ., when dosing or filling vessels .
Stability signal
High accuracy
Medium
accuracy*
Fast
* Factory setting