Center the calibration weight on the weighing pan.
An acoustic signal will indicate the end of the
calibration process.
SENSITIVITY TEST (special function for balances
with a built-in calibration weight)*
Rather substantial changes in barometric pressure
and temperature may affect the display response
of these highly accurate balances. To ensure that you
obtain the full accuracy of these balances, even
when you use the entire weighing range, we have
implemented a sensitivity test function. It will help
you decide quickly whether or not you need to
recalibrate your balance (for example, to maintain the
same accuracy during long-term weighing series).
Press the tare key
(10)
for at least 2 seconds until
“
C–I
” and “
CAL
” are displayed (next to the
f
key).
Select the
sensitivity test
by pressing the
F
key
(12) twice
.
* = Not on verified LC balances
Important Note for Accuracy Class
K
Verified Balances/Scales Approved for Use
as Legal Measuring Instruments in the EU:
After verification, the “external calibration” function
is blocked by the access switch
(5)
(adjusted
to the left and secured against alteration of the
switch setting).
1–23
2 sec.