Weighing in Percent with Additional Requirements
Changing the Reference Percentage
You can change the reference percentage to meet your individual requirements.
The reference percentage setting will be displayed above the unit symbol after you
press
c
.
To select the change function:
Hold down
i
for more than 2 seconds
To change the setting:
Use
f
identified by – ^ –
If you enter a wrong number, press
c
to clear it.
To store the percentage:
– temporarily until you turn off the power: briefly
press
F
– p
ermanently after you turn off the power (in the non-
volatile memory): press
F
for more than 2 seconds
Note: This setting is not canceled by the reset code 9 – –1
o
.
Changing the reference percentage
Code
No change
2 3 1
In cycles: 5,10, 20, 50,100, 5,10, …
* 2 3 2
In increments of one
2 3 3
Storage Parameter for the Reference Weight
The reference weight is stored
Code
with full accuracy according to the internal resolution
3 5 1
according to the display accuracy
* 3 5 2
Display Parameter for Readouts in Percent
The readout in percent is displayed
Code
without a decimal place
3 6 1
with one decimal place
* 3 6 2
with two decimal places
3 6 3
with three decimal places
3 6 4
Use in Combination with Other Programs
Program
Code
Tare memory
2 2 2
Over/under checkweighing (net weight)
2 2 3
Over/under checkweighing (difference in weight)
2 2 4
* = factory setting
4–8