Stability Symbol Delay
This setting allows your balance to compensate for individual interfering factors
which slowly subside, such as turbulent air currents generated within the weighing
chamber of an analytical balance. You can therefore choose to have the stability
symbol displayed later.
Code
No delay
1 4 1
Short delay
* 1 4 2
Long delay
1 4 3
Extremely long delay
1 4 4
Tare Parameter
You can define when the balance will perform the taring operation:
Code
At any time
** 1 5 1
Not until the readout is stable
* 1 5 2
Only at stability
1 5 3
Auto Zero Function
When this zero tracking function is activated, any changes off the zero readout
that are equal to a defined fraction of digits per second are automatically tared.
In other words, it ensures a stable zero.
Code
Auto Zero on
* 1 6 1
Auto Zero off
1 6 2
* = factory setting
** = not applicable to verified balances approved for use as legal
measuring instruments
2–6