7. If the instrument does not switch back to the uncovered
state, try with a lower sensitivity value.
Select the sensitivity value with the help of Table 4.
SENSITIVITY
DIELECTRIC CONSTANT
1
>
30
2
20-30
3
5-20
4
2-5
5
>
1.5 &
<
2
Table 4: Sensitivity
14
Failsafe
In a condition of device failure, known errors and input
power failure the outputs of the device resemble the ALARM
condition. This is meant to prevent overflow or dry run con-
ditions in case of failures.
Prevent Overflow - High Level Switch
: Failsafe High
(default) is set by moving the FS 1 and FS 2 switch for RE-
LAY 1 and 2 to OPEN position.
1. When not in contact with the material, LED turns
GREEN.
2. When in contact with the material, LED turns RED.
ENTER
COAT-ENDURE
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 26: Failsafe High
Prevent Dry run - Low Level Switch
:
Note:
LO in top cover indicates failsafe low.
Failsafe Low is set by moving the FS 1 and FS 2 switch for
RELAY 1 and 2 switch to CLOSE position.
1. When in contact with the material, LED turns GREEN.
2. When not in contact with the material, LED turns RED.
15
Display Indications
As seen in Fig.28, Coat-Endure has STATUS LEDs and
a 6-point LED bar display.
15.1
STATUS LED
Two separate STATUS LED sections indicate switching
status for RELAY 1 and RELAY 2.
ENTER
COAT-ENDURE
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 27: Failsafe Low
ENTER
COAT-ENDURE
+
-
NC
C
NO
ON
DIP
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
Lo
+
-
OUT
L
N
E
Relay-1
PNP
STATUS
LED
Figure 28: LED Indication on Top Cover
•
RED
LED: Alarm Condition
•
GREEN
LED: Normal Condition
Continuous blinking of Red or Green LED could indicate an
error, refer to Error Indication
15.2
LED Bar Display
A numeric LED Bar displays helps to visualize swithching
point and senstivity values.
Use of LED Bar Display
: The bar display can be used
for two purpose:
•
Sensitivity: Refer to section Sensitivity for more details.
•
Display Mode: The display mode shows the material
distance from the instrument’s probe. The movement
of LEDs in upward direction depicts the movement of
application media towards Coat-Endure’s Probe and
vice-versa, as shown in Figure 29.
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
Upward movement of LEDs
Downward movement of LEDs
Material get closer
Material moves far
Figure 29: Demonstration of LED Bar Display
16
Factory Reset
Follow the steps given below to reset the time delays and
sensitivity value to default values. Factory Rest does not
reset any calibration values:
1. Set the CALIBRATE, COVER and UNCOVER switches
to CLOSE position.
Sapcon Instruments Pvt.Ltd.