Coat-Endure
User Manual & Datasheet V 3.0
ENTER
COAT-ENDURE
ON
DIP
1
2
3
4
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 13: High Calibration Switch Position
•
Press and hold ENTER key, the STATUS LED for RE-
LAY 1 will glow in RED color as shown in Figure 14.
ENTER
COAT-ENDURE
ON
DIP
1
2
3
4
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 14: Setting Calibration
•
Release the ENTER key and set the HI CALIBRATE
switch back to OPEN position as shown in Figure 15.
ENTER
COAT-ENDURE
ON
DIP
1
2
3
4
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 15: Saving Calibration
•
High Calibration is complete.
11
Cover Delay
Note:
Set the value of Cover Delay between 1-25 secs.
When the application material covers the probe, the
changeover of the output can be delayed by a pre-
determined time. This time is called Cover Delay. For a
different value of cover delay, the number of blinks can be
adjusted as per requirement. Follow the procedure given
below for setting cover delay:
1. Ensure that all DIP switches are in OPEN position as
shown in Figure 16. Make sure that STATUS LED is
not blinking for error.
2. Set the COVER switch to CLOSE (Opposite of OPEN
in a DIP switch) position as shown in Figure 17.
3. Press ENTER. The STATUS LED for RELAY 1 starts
blinking. Blink the STATUS LED according to value of
cover delay.
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI
CALIBRATE
FS1
FS2
CLOSE
Figure 16: DIP Switch
ENTER
COAT-ENDURE
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 17: Cover Switch Position
ENTER
COAT-ENDURE
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 18: Setting Cover Delay
4. Cover delay is entered, but not saved. To save and test
the cover delay, set the COVER switch back to OPEN
position as shown in Figure 19.
ENTER
COAT-ENDURE
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 19: Saving Cover Delay
5. To test, dip Coat-Endure into the application material
until the switching point is reached.
6. The STATUS LED will start blinking RED if the switch
point is reached. It will blink for the number of seconds
for which the cover delay is set. 1 blink is equal to 1
second during switching.
12
Uncover Delay
Note:
Set the value of Uncover Delay between 1-25
secs.
When the application material uncovers Coat-Endure’s
probe, the changeover of the output can be delayed by a
pre-determined time. This time is called Uncover Delay. For
a different value of uncover delay, the number of blinks can
be adjusted as per requirement. Follow the below proce-
dure for setting uncover delay:
Sapcon Instruments Pvt.Ltd.
R
10