‑ 7 ‑
6‑3. Zero adjustment
Press the probe against the substrate for adjustment (zero plate).
[
Indication on the LCD
⑦
changes as follows.]
↓
←
The buzzer emits a beeping
sound.
Measured value [XX] is
Zero plate
indicated on the LCD
⑦
.
↓
Press the
「
ZERO
」
key
②
.
(
Both the probe contact to zero plate or not is possible.)
↓
←
The buzzer emits a beeping sound.
[
CAL
]
is indicated on the upper left of the
LCD
⑦,
[
0
]
is indicated.
↓
←
The buzzer emits a beeping sound.
Press the probe against the zero plate several times, when the [0] or closed values are indicated
on the LCD
⑦
, the adjustment is acceptable.
When the indications are away from [0], repeat above procedures several times.
※When 「LLL」is indicated while zero adjusting, its adjusting point is widely deviated
from a right position.
Repeat the zero adjustment 2 to 4 times after confirming no coating is applied to the
substrate and confirm zero is stably indicated.
XX
μm
Non‑Fe
CAL
0
Non‑Fe
0
μm
Non‑Fe