background image

‑ 7 ‑

6‑3. Zero adjustment

Press the probe against the substrate for adjustment (zero plate).

Indication on the LCD

changes as follows.]

The buzzer emits a beeping

sound.

Measured value [XX] is

Zero plate

indicated on the LCD

.

Press the

ZERO

key

.

(

Both the probe contact to zero plate or not is possible.)

The buzzer emits a beeping sound.

[

CAL

]

is indicated on the upper left of the

LCD

⑦,

[

]

is indicated.

The buzzer emits a beeping sound.

Press the probe against the zero plate several times, when the [0] or closed values are indicated
on the LCD

, the adjustment is acceptable.

When the indications are away from [0], repeat above procedures several times.

※When 「LLL」is indicated while zero adjusting, its adjusting point is widely deviated

from a right position.
Repeat the zero adjustment 2 to 4 times after confirming no coating is applied to the
substrate and confirm zero is stably indicated.

XX

μm

Non‑Fe

CAL

Non‑Fe

μm

Non‑Fe

Summary of Contents for EDY-5000

Page 1: ...r read this INSTRUCTION MANUAL thoroughly and use the Meter correctly Keep this INSTRUCTION MANUAL carefully and refer to this when necessary In the event of any doubt arising the original INSTRUCTION...

Page 2: ...Disconnecting the probe 5 OPERATING INSTRUCTIONS 6 6 1 Power ON 6 6 2 Power OFF 6 6 3 Zero adjustment 7 6 4 Standard calibration CAL 8 6 5 Resetting the working curve 9 6 6 Measurement 10 ADDITIONAL F...

Page 3: ...to measure the thickness D of non conductive film on non magnetic metallic substrate APPLICATIONS The Meter is used for non destructive measurement of the thickness of a non conductive film e g ALMIT...

Page 4: ...CAL 10 m to 5 00 mm Probe One point contact constant pressure type with V groove 18 55mm Additional function 1 Conversion of key lock mode 2 Auto power off about 5 minutes Power source Dry batteries...

Page 5: ...ON OFF key key Key for standard calibration with the standard thickness plate Printer connector Connector for connecting the printer Probe connector Connector for connecting the probe LCD display Indi...

Page 6: ...ust dirt or dust if any 5 2 Connecting Disconnecting the probe Carry out the connection disconnection of the probe by the following procedures Be sure to keep the Power OFF The probe cannot be substit...

Page 7: ...he buzzer emits a beeping sound The Meter is ready for measurement When power ON the last measured value of the previous measurement is indicated When the first power ON or no measured data is in the...

Page 8: ...he upper left of the LCD is indicated The buzzer emits a beeping sound Press the probe against the zero plate several times when the 0 or closed values are indicated on the LCD the adjustment is accep...

Page 9: ...s ready for measurement Press the probe several times against the standard thickness plate placed on the zero plate When the indication on the LCD is the same thickness or near with the standard thick...

Page 10: ...CD and is indicated blinking The buzzer emits a beeping sound for every press of the key The buzzer emits beeping sounds E is indicated for 3 seconds The Meter is restored to the possible condition of...

Page 11: ...easuring condition is restored by working of the resuming function at the succeeding power ON Correct usage Incorrect usage Hold the mid part of the probe and Don t tilt the probe Don t slide the prob...

Page 12: ...teries with fresh specified dry batteries Battery Dry battery R03 1 5V 4 pcs MAINTENANCE AND INSPECTION Use the Meter within the range of 0 to 40 Avoid to expose the Meter to dew condensation wet with...

Page 13: ...The measurement of the film having a metal substance causes a measurement error With respect to an elastic film place a standard thickness plate with 30 to 50 m on it and subtract the thickness of the...

Page 14: ...ector Iron piece detector Manufacturer Sanko Electronic Laboratory Co Ltd 2 6 4 Tokyo branch Shibata Bldg Uchikanda Chiyoda ku Tokyo Japan 101 0047 81 3 3254 5031 81 3 3254 5038 Tel Fax Osaka branch K...

Reviews: