3), asked in detail about the patient's medical history, including received what forms of treatment
previously to determine whether he / she is suitable to accept the 808nm laser hair removal treatment.
4), find out why patients seek treatment and his / her expectations.
5) discuss the 808nm laser hair removed treatment plan with the patient.
7.4.2 explanation
Newly diagnosed, the attending physician should tell the patient the following(or other qualified medical
personnel):
• treatment, there may be some discomfort / pain;
• After treatment, there may be temporary swelling / edema occur;
• To achieve the desired therapeutic effect may require several treatment, usually 4-6 times. Efficacy of
the treatment need to follow the gradual expression and the patient's actual expectation;
• There have temporary hyperpigmentation and skin texture changes and other minor side effects risk.
7.4.3 eye protection
When using 808nm diode laser hair removal machine, all persons in the treatment room (including
patients and all health care workers) must wear protective glasses to protect the retina. When the intense
pulsed light is used near the eye,doctors should give the patient over their eyes. Of course, the best
protective measures is to let the patient closed his eyes before the laser pulse launching after the system
ready to issue a "beep" sound,.
7.4.4, photo
Photographic archive for the patient before and after surgery (the patient face left, right and front).
Because many patients are unable to objectively evaluate treatment progress, these photos can be used
as evidence.
Patient photography should be used in the same environment similar to the speed, flash index and the
focal length, so the photos can do different time objective comparison.
7.5 Treatment
1) • allow patients to wear protective glasses sitting in the chair. Operators also wear safety glasses, and
remind the patient do not open their eyes during surgery.Operator select the best operating position
between the patients’ face and touch control panel.
2) • General from inconspicuous parts of the face, such as the front of the ear begin treatment. In the
treated area coated with 1-2mm cold gel, which will help cool the skin under the laser pulse and increase
the penetration of light pulses on the skin. Pulse emission in the cold gel, each light pulse treatment to
receive the site should be coated with a new cooling gel.
3) • placed in the treatment of head and perpendicular to the cold 1-2mm thick gel, the optical crystal
contacts in the skin, rather than the cold gel
4) • When the system issued a "tick" and ready sound. Pressing the trigger switch, a laser pulse.
5) (Selection of parameter is a very important aspect in clinical training, We recommend doctors firstly