SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11)
Page 65 of 79
Chapter 9. Troubleshooting
Property damage may occur due to improper repair.
If personnel other than authorized by the manufacturer carries out the repair, it may irreversibly damage the
device.
‐
Repair should only be performed by the manufacturer.
9.1. Troubleshooting
Issue
Recommended action(s)
The hip joint of the device becomes
stuck during operation.
•
If the actuator becomes stuck during operation, do not apply excessive
force to the device.
•
First, turn off and then take off the device.
•
Check what the problem is while gently moving the corresponding joint.
•
If the problem persists or occurs repeatedly, please contact the
manufacturer.
Case / Frame / Fastener breaks
•
If the hardware is damaged, do not operate the device. Contact the
manufacturer.
There is unusual noise coming from
the device.
•
If you hear noise from the device that is not associated with normal use,
please contact the manufacturer.
No Bluetooth connection.
•
Make sure that Bluetooth is activated on the tablet PC. (Check this from
Quick Settings on the top of the tablet screen).
•
Restart the GEMS‐H App. (Press the button left of the home button on
the bottom of the tablet screen to completely close the app).
•
Restart the GEMS‐H Device. (Use the power switch located at the bottom
of the control pack).
•
Restart the associated tablet PC.
Assistance is uncomfortable or the
patient cannot feel the assistive
force.
•
Make sure the LED indicator is not indicating a “STOP” state.
•
Repeat the procedure for resetting the Stand Pose to zero when the
patient is stationary and standing upright. This procedure may be found
in section 6.8.3, under “Mitigation of risks due to excessive joint angle
movement.”
•
Restart the assistance algorithm with default control parameter values
(or quit the GEMS‐H and run it again).
•
Check whether the hip angle/torque value during walking is normal or
not.
Summary of Contents for GEMS-H
Page 1: ...GEMS H Instructions for Use ...
Page 19: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 19 of 79 4 6 Marking plate ...
Page 52: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 52 of 79 6 7 3 3 Example of parameter control ...
Page 54: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 54 of 79 Error message examples ...
Page 64: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 64 of 79 ...
Page 67: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 67 of 79 ...