8.2. Regular maintenance (Weekly Checklist)
Check that the bolts on the actuator cover are not loose.
Check that the bolts on the battery cover are not loose.
Check that the bolts on the thigh support frame are not loose.
8.3. Cleaning
Risk of electrical shock hazard due to moisture!
Electrical shock may be delivered to the patient and the user due to excessive moisture when the device is
turned on while cleaning.
‐
Do not clean while in use.
‐
Be sure to turn off the power before cleaning.
Property damage may occur due to improper cleaning procedures or agents!
The product may be damaged if it is cleaned and disinfected with improper procedures or agents.
‐
Use only this procedure for cleaning.
If the device is used by more than one patient, it is essential to clean the device with disinfectant before use.
To prevent cross‐infection between patients, the device (i.e., waist part, thigh support frames, and thigh support
straps) should be cleaned after each use. The device should be disinfected as follows:
①
Dampen (but do not soak) a lint‐free cloth with purified water.
②
Wipe the surface of the device for at least five (5) minutes.
③
Wipe the surface of the device with an (isopropyl) alcohol‐based disinfectant for at least five (5) minutes.
‐ Recommended Disinfectant: Cavi‐Wipes (Metrex Research, Kerr) or equivalent
Summary of Contents for GEMS-H
Page 1: ...GEMS H Instructions for Use ...
Page 19: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 19 of 79 4 6 Marking plate ...
Page 52: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 52 of 79 6 7 3 3 Example of parameter control ...
Page 54: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 54 of 79 Error message examples ...
Page 64: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 64 of 79 ...
Page 67: ...SAIT IFU GEMS H Rev 1 1 2019 11 Page 67 of 79 ...