[25]
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
7.5
Immunity and Compliance Level
Immunity test
IEC 60601 Test Level
Actual Immunity Level
Compliance Level
Radiated RF
IEC 61000-4-3
10V/m
80MHz - 2.7GHz
80 % AM at 1 kHz
10V/m
10V/m