![Samsung SDS S-PATCH3-Cardio User Manual Download Page 22](http://html1.mh-extra.com/html/samsung-sds/s-patch3-cardio/s-patch3-cardio_user-manual_3754107022.webp)
[22]
7.
Information on EMC
7.1
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
-
The EUT is intended for use in the electromagnetic environment specified below. The
customer or the user of the EUT should assure that it is used in such an environment.
Immunity test
Compliance
Electromagnetic environment - Guidance
RF Emissions
CISPR 11
Group 1
The S-PATCH3 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment
RF Emissions
CISPR 11
Class B
The S-PATCH3 is suitable for use in ail establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes
7.2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
-
The EUT is intended for use in the electromagnetic environment specified below. The
customer or the user of the EUT should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8kV Contact
±2,4,8,15 kV air
±8kV Contact
±2,4,8,15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.