Tel: 886. 909 602 109 Email: [email protected]
www.salukitec.com
106
4. Amplitude Accuracy and Flatness
Sensitivity
Amplitude
Lower Limit
Reading
Upper
1V
1.0000Vrms
0.9800V
1.0200V
200mV
200.00Vrms
198.99mV
202.00mV
100mV
100.00Vrms
98.00mV
102.00mV
20mV
20.000Vrms
19.60mV
20.400mV
10mV
10.000Vrms
9.800mV
10.200mV
Sensitivity
Amplitude
Lower Limit
Reading
Upper
200mV
24kHz
196mV
204mV
200mV
48kHz
196mV
204mV
200mV
72kHz
196mV
204mV
200mV
96kHz
196mV
204mV
5. Amplitude Linearity
Sensitivity
Amplitude
Lower Limit
Reading
Upper Limit
1V
1.0000Vrms
0.9900V
1.0100V
100.00Vrms
0.0990V
0.1010V
10.000Vrms
0.0099V
0.0101V
6. Frequency Accuracy
Frequency
Lower Limit
Reading
Upper Limit
10kHz
9.990kHz
10.010kHz
7. SINE OUT Amplitude and Flatness
Sensitivity
SINE OUT Ampl.
Lower Limit
Reading
Upper Limit
1V
2.000Vpp
(0.7071Vrms)
0.6859V
0.7283V
200mV
0.400Vpp
(141.42Vrms)
137.18mV
145.66mV
50mV
0.100Vpp
(35.355Vrms)
34.294mV
36.416mV
10mV
0.020Vpp
(7.071Vrms)
6.859mV
7.283mV
SINE OUT
Frequency
Lower Limit
Reading
Upper Limit
2.000Vpp
24 kHz
0.6859V
0.7283V
(0.7071Vrms)
48 kHz
0.6859V
0.7283V
72 kHz
0.6859V
0.7283V