and the expand is 10
8
V/A (i=3).
IGND(?){i}
The IGND command sets or queries the input shield grounding. The parameter i selects
Float (i=0) when the input ground is connected to the instrument ground with a 10 kΩ
resistor or Ground (i=1) when the input ground is connected to the instrument ground with
a 10 Ω resistor
ICPL(?){i}
The ICPL command sets and queries the input coupling. The parameter i selects AC (i=0)
or DC (i=1).
ILIN(?){i}
The ILIN command sets and queries the notch filter status. The parameter i selects to close
notch filters (i=0), open the 50Hz notch filter (i=1), open the 50Hz and 100Hz notch filter
at the same time (I=2) or open the 100Hz notch filter (i=3).
4.2.4 Sensitivity and Time Constants Commands
SENS(?){i}
The SENS command sets or queries the sensitivity. The parameter i is used to select a
different range.
i sensitivity
i sensitivity
1: 1 nV/fA
15: 50 uV/pA
2: 2 nV/fA
16: 100 uV/pA
3: 5 nV/fA
17: 200 uV/pA
4: 10 nV/fA
18: 500 uV/pA
5: 20 nV/fA
19: 1 mV/nA
6: 50 nV/fA
20: 2 mV/nA
7: 100 nV/fA
21: 5 mV/nA
8: 200 nV/fA
22: 10 mV/nA
9: 500 nV/fA
23: 20 mV/nA
10: 1 uV/pA
24: 50 mV/nA
11: 2 uV/pA
25: 100 mV/nA
12: 5 uV/pA
26: 200 mV/nA
13: 10 uV/pA
27: 500 mV/nA
14: 20 uV/pA
28: 1 V/uA
RMOD(?){i}
The RMOD command sets or queries the mode of dynamic reserve. The parameter i selects
<Low Noise> (i=0), <Normal> (i=1) or <High Reserve> (i=2).
OFLT(?){i}
The OFLT command sets or queries the time constant of filters. The parameter i selects a
time constant below.