Options
Settings
Comments
Slope
Rising
Falling
Select the slope type of signal.
Mode
Auto
Normal
Select the type of trigger. The Normal mode is best for most pulse width trigger
applications.
Coupling
AC
DC
Noise Reject
HF Reject
LF Reject
Selects the components of the trigger signal applied to the trigger circuitry.
Next Page
Vertical
V1
V2
Adjust the vertical window by setting two trigger levels. Select this option and press
F3 to choose V1 or V2.
When
=
≠
<
>
Select the trigger condition.
Time
20ns to 10.0sec
With this option highlighted by pressing F4, turn the multi-functional knob to set the
time span.
8.
Swap Trigger:
As a feature of analog oscilloscopes, it gives stable displays of signals at two different frequencies. Mainly it uses a
specific frequency to switch between two analog channels CH1 and CH2 so that the channels will generate swap
trigger signals through the trigger circuitry.
Options
Settings
Comments
Swap Trigger
Mode
Auto
Normal
Select the type of trigger.
Channel
CH1
CH2
Push an option such as CH1, select the channel trigger type and set the menu interface.
Below list options in submenus. Swap Trigger allows CH1 and CH2 to select different trigger modes and to display waveforms on a same screen. That
is, both channels can choose the following four trigger modes.
Type
Edge
Slope
Rising
Falling
Coupling
AC
Push F3 or F4 to select the components of the trigger signal applied to the trigger circuitry.