background image

 

PRO

check 

®    

 

Ultrasonic Oxygen Indicator

 

 

 

Trouble Shooting Checklist

 

 

Complaint

 

Probable Cause

 

Suggested 

Corrective Action

 

Display is 

blank 

when 

unit is turned on. 

Bad batteries. 
Corroded battery 
terminals. 

Replace batteries. 
Clean battery terminals. 

When unit is turned on 
display flashes 

"PRO O2 @ CAL"

 

and then turns off. 

 

Unit requires 
calibration. 

Perform "Full 
Calibration Procedure" 
outlined in Operation 
Manual. 

 

 

 

Display reads 

91.8% 

on 

a source of U.S.P. 
oxygen (> 99.0%). 

 

Unit is operating in 
mode to check 
concentrators and is 
adjusting for 
concentrator's output 
of argon. 

With unit still connected to 
source of U.S.P. oxygen (> 
99.0%) push 
and continue  to hold 
down 

pure 

O

2

/calibrate

” 

button to 

temporarily remove argon 
correction from unit's 

 

 

 

With unit still connected to 
source of U.S.P. oxygen  (> 

Display reads between 

98% to 102% 

(but not 

99.9%) on a source of 
U.S.P. oxygen 
(> 99.0%) with button 
held down. 

 

 

Unit requires quick 
calibration. 

99.0%) push and hold 
down 

pure O

2

calibrate

” 

button until 

display reads 99.9% - 101 
(approximately 20 
seconds). The unit is then 
recalibrated. 

Unit is out of 98% to 
102% range after quick 
calibration. 

Unit requires full 
calibration. 

Perform "Full 
Calibration Procedure" 
outlined in User Manual. 

"---" 

appears on display 

while unit is being 
calibrated. 

 

Incomplete 
calibration. 

Release the 

"

pure 

O /calibrate

button 

2 

and restart calibration 
process. 

 

Display shows 

"-U-"

Indicates an out of 
range reading. 

Perform "Full Calibration 
Procedure" outlined in 
User Manual. 

 

 

 

 

13 

Summary of Contents for RO2 check

Page 1: ...J CAUTION 30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024 Email customercare salterlabs com www salterlabs com...

Page 2: ...30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024...

Page 3: ...Liquid Oxygen Vessels or Oxygen Cylinders 5 VI Verification Calibration Instructions 6 Pure Oxygen Verification Procedure 6 Quick One Step Pure Oxygen Calibration 6 Full Two Step Calibration Procedure...

Page 4: ...tal device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and...

Page 5: ...ff button will turn the unit on or off 2 The pure O2 calibrate button when pressed and held down while connected to a compressed gas or liquid U S P oxygen 99 0 source verifies that the unit is proper...

Page 6: ...ncentration measurement mode can be quickly and easily verified by following the Pure Oxygen Verification procedure on page 6 of this manual 3 If accreditation organizations or internal policies and p...

Page 7: ...2 check device off press and release the on off button The unit will display OFF and shut down To conserve battery life it automatically shuts down if any one of the unit s functions is not utilized f...

Page 8: ...O2 check unit on 4 When unit stabilizes the unit display will read Concentrator O2 91 8 approx concentration 5 Press and continue to hold down the pure O calibrate button 6 Unit will display Pure O2 a...

Page 9: ...l display Pure O2 and a numeric reading 7 If the unit is properly calibrated the display should read between 99 and 101 The PRO2 check is ready to use 8 If the unit display reads err U or any other re...

Page 10: ...elease Calibration Pure O2 button 6 You will see three dashed lines scroll in circles Then the scrolling will stop and then you will see Cal End and read 91 7 Note Reading may range from 91 to 92 This...

Page 11: ...ted on the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by slidi...

Page 12: ...ease verify calibration if used over 5 000 feet Note If readings do not seem accurate recalibrate the unit Vibration It is recommended that the unit be used in a stable position to prevent the display...

Page 13: ...hargeable type battery is not recommended Operating Temperature 32 F to 105 F 0 C to 41 C Storage Temperature 29 F to 160 F 34 C to 71 C Specifications Concentration Mode Measured Oxygen Concentration...

Page 14: ...ure that the unit is holding calibration against a U S P oxygen 99 0 source the unit does not require calibration My accreditation organization as well as our internal policies require periodic calibr...

Page 15: ...pprox What s wrong Nothing When in the Concentrator Mode the electronic components of the indicator automatically compensates for the argon found in the gas produced by an oxygen concentrator As a res...

Page 16: ...nd continue to hold down pure O2 calibrate button to temporarily remove argon correction from unit s With unit still connected to source of U S P oxygen Display reads between 98 to 102 but not 99 9 on...

Page 17: ...n Manual Display shows err during use Unit has made a reading outside normal ranges Check calibration Make sure gas being measured is from oxygen concentrator LO BATT appears in display screen when un...

Page 18: ...be responsible for all shipping charges to return the unit to Salter Labs Salter Labs will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided...

Reviews: