Ultrasonic Oxygen Indicator

 

2

 

 

 

 

INSTRUCTIONS FOR SPOT CHECKING 

LIQUID OXYGEN VESSELS or OXYGEN CYLINDERS

 

 

Although the 

PRO

check

® 

Ultrasonic Oxygen Indicator 

is designed, as its 

name implies, to check the various functions of an oxygen concentrator, it 
can also be used to perform spot checks (± 2%) of gaseous or liquid 
oxygen. 

 

To utilize the 

PRO

check

® 

device for this purpose follow this procedure: 

 

1.

 

Connect the inlet port on the 

PRO 

check

® 

unit to a gaseous or 

liquid  oxygen source (cylinder or liquid oxygen) via a supply tube.

 

 

2.

 

Turn the oxygen flow on to 2 LPM.

 

 

3.

 

Press the 

on/off

” 

button to turn the 

PRO

check

® 

unit on.

 

 

4.

 

When unit stabilizes the unit display will read "

Concentrator O

2

 

91.8" 

(approx concentration.)

 

5.

 

Press and continue to hold down the 

pure O /calibrate

” 

button.

 

 

6.

 

Unit will display 

"Pure O2" 

and a numeric reading.

 

 

7.

 

If the source gas is oxygen, the display should read between 

99% 

and 101%

.

 

 

8.

 

If the unit display reads 

"err"

"---", "-U-" 

or any other reading 

other than 99.9 (± 2%), the source gas may not be pure oxygen and 
should be quarantined for retesting using a more sophisticated method 
such as a Servomex* oxygen analyzer.

 

 

NOTE: 

PRO 

check

® 

Ultrasonic Oxygen Indicator 

can be used to spot 

check  the identity and purity of gaseous or liquid oxygen only, it cannot be 
used to  perform or verify U.S.P. oxygen (> 99.0%) testing since it cannot 
provide  the required U.S.P. oxygen (> 99.0%) accuracy of ± 0.1%. 

 

 

* Servomex Inc. - Sugarland, TX

 

 

 

 

 

 

 

5

 

Summary of Contents for RO2 check

Page 1: ...J CAUTION 30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024 Email customercare salterlabs com www salterlabs com...

Page 2: ...30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024...

Page 3: ...Liquid Oxygen Vessels or Oxygen Cylinders 5 VI Verification Calibration Instructions 6 Pure Oxygen Verification Procedure 6 Quick One Step Pure Oxygen Calibration 6 Full Two Step Calibration Procedure...

Page 4: ...tal device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and...

Page 5: ...ff button will turn the unit on or off 2 The pure O2 calibrate button when pressed and held down while connected to a compressed gas or liquid U S P oxygen 99 0 source verifies that the unit is proper...

Page 6: ...ncentration measurement mode can be quickly and easily verified by following the Pure Oxygen Verification procedure on page 6 of this manual 3 If accreditation organizations or internal policies and p...

Page 7: ...2 check device off press and release the on off button The unit will display OFF and shut down To conserve battery life it automatically shuts down if any one of the unit s functions is not utilized f...

Page 8: ...O2 check unit on 4 When unit stabilizes the unit display will read Concentrator O2 91 8 approx concentration 5 Press and continue to hold down the pure O calibrate button 6 Unit will display Pure O2 a...

Page 9: ...l display Pure O2 and a numeric reading 7 If the unit is properly calibrated the display should read between 99 and 101 The PRO2 check is ready to use 8 If the unit display reads err U or any other re...

Page 10: ...elease Calibration Pure O2 button 6 You will see three dashed lines scroll in circles Then the scrolling will stop and then you will see Cal End and read 91 7 Note Reading may range from 91 to 92 This...

Page 11: ...ted on the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by slidi...

Page 12: ...ease verify calibration if used over 5 000 feet Note If readings do not seem accurate recalibrate the unit Vibration It is recommended that the unit be used in a stable position to prevent the display...

Page 13: ...hargeable type battery is not recommended Operating Temperature 32 F to 105 F 0 C to 41 C Storage Temperature 29 F to 160 F 34 C to 71 C Specifications Concentration Mode Measured Oxygen Concentration...

Page 14: ...ure that the unit is holding calibration against a U S P oxygen 99 0 source the unit does not require calibration My accreditation organization as well as our internal policies require periodic calibr...

Page 15: ...pprox What s wrong Nothing When in the Concentrator Mode the electronic components of the indicator automatically compensates for the argon found in the gas produced by an oxygen concentrator As a res...

Page 16: ...nd continue to hold down pure O2 calibrate button to temporarily remove argon correction from unit s With unit still connected to source of U S P oxygen Display reads between 98 to 102 but not 99 9 on...

Page 17: ...n Manual Display shows err during use Unit has made a reading outside normal ranges Check calibration Make sure gas being measured is from oxygen concentrator LO BATT appears in display screen when un...

Page 18: ...be responsible for all shipping charges to return the unit to Salter Labs Salter Labs will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided...

Reviews: