Ultrasonic Oxygen Indicator

 

 

 

 

WARNING NOTES

 

 

USAGE: 

The 

PRO

check

® 

Ultrasonic  Oxygen  Indicator  is  designed  for 

intermittent  usage  to  determine  the 

oxygen  concentration 

produced  by  an 

oxygen  concentrator.  It  may  also  be  utilized  to  do  spot checks  of  gaseous or 
liquid  oxygen.  This  instrument  has  no  clinical  diagnostic  or  therapeutic 
functions. 

 

WARNING: 

The  unit  is  not  designed  nor  intended  for  use  in  anesthesia 

applications,  or  for  monitoring  oxygen  concentration  from  any  source  other 

than a conventional oxygen concentrator using molecular sieve beds. This unit 

will not withstand fluid spillage. The unit will not withstand mechanical shock 

or vibration. 

 

NOTE: 

This equipment  has  been tested and  found to comply  with the limits 

for  a  Class  B  digital  device,  pursuant  to  Part  15  of  the  FCC  Rules.  These 
limits  are  designed  to provide reasonable  protection  against  harmful 
interference  in  a  residential  installation.  This  equipment  generates,  uses,  and 
can radiate radio frequency energy and if not installed in accordance with the 
instructions,  may  cause  harmful  interference  to  radio  communications. 
However, there is no guarantee that interference  will not occur in a particular 
installation.  If  this  equipment  does  cause  interference  to  radio  or  television 
reception,  which can be determined by turning the equipment off and on, the 
user  is  encouraged  to  try  and  correct  the  interference  by  one  or  more  of  the 
following measures: 

 

 

       

Reorient or relocate the receiving antenna

 

      

Increase the distance between the equipment and receiver.

 

      

Connect the equipment into an electrical outlet on a different 

circuit from which the receiver is connected.

 

      

Consult the dealer or an experienced radio/TV technician for help.

 

 

 

 

 

 

 

 

 

Summary of Contents for RO2 check

Page 1: ...J CAUTION 30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024 Email customercare salterlabs com www salterlabs com...

Page 2: ...30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024...

Page 3: ...Liquid Oxygen Vessels or Oxygen Cylinders 5 VI Verification Calibration Instructions 6 Pure Oxygen Verification Procedure 6 Quick One Step Pure Oxygen Calibration 6 Full Two Step Calibration Procedure...

Page 4: ...tal device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and...

Page 5: ...ff button will turn the unit on or off 2 The pure O2 calibrate button when pressed and held down while connected to a compressed gas or liquid U S P oxygen 99 0 source verifies that the unit is proper...

Page 6: ...ncentration measurement mode can be quickly and easily verified by following the Pure Oxygen Verification procedure on page 6 of this manual 3 If accreditation organizations or internal policies and p...

Page 7: ...2 check device off press and release the on off button The unit will display OFF and shut down To conserve battery life it automatically shuts down if any one of the unit s functions is not utilized f...

Page 8: ...O2 check unit on 4 When unit stabilizes the unit display will read Concentrator O2 91 8 approx concentration 5 Press and continue to hold down the pure O calibrate button 6 Unit will display Pure O2 a...

Page 9: ...l display Pure O2 and a numeric reading 7 If the unit is properly calibrated the display should read between 99 and 101 The PRO2 check is ready to use 8 If the unit display reads err U or any other re...

Page 10: ...elease Calibration Pure O2 button 6 You will see three dashed lines scroll in circles Then the scrolling will stop and then you will see Cal End and read 91 7 Note Reading may range from 91 to 92 This...

Page 11: ...ted on the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by slidi...

Page 12: ...ease verify calibration if used over 5 000 feet Note If readings do not seem accurate recalibrate the unit Vibration It is recommended that the unit be used in a stable position to prevent the display...

Page 13: ...hargeable type battery is not recommended Operating Temperature 32 F to 105 F 0 C to 41 C Storage Temperature 29 F to 160 F 34 C to 71 C Specifications Concentration Mode Measured Oxygen Concentration...

Page 14: ...ure that the unit is holding calibration against a U S P oxygen 99 0 source the unit does not require calibration My accreditation organization as well as our internal policies require periodic calibr...

Page 15: ...pprox What s wrong Nothing When in the Concentrator Mode the electronic components of the indicator automatically compensates for the argon found in the gas produced by an oxygen concentrator As a res...

Page 16: ...nd continue to hold down pure O2 calibrate button to temporarily remove argon correction from unit s With unit still connected to source of U S P oxygen Display reads between 98 to 102 but not 99 9 on...

Page 17: ...n Manual Display shows err during use Unit has made a reading outside normal ranges Check calibration Make sure gas being measured is from oxygen concentrator LO BATT appears in display screen when un...

Page 18: ...be responsible for all shipping charges to return the unit to Salter Labs Salter Labs will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided...