IntuBrite VLS 6600 Series Edge Video Laryngoscope Operating 

Instructions 

Intended Use: 

The rigid laryngoscope is intended for use in 

laryngoscopy to examine and visualize the upper airway and glottis. 
Please read the instructions carefully before attempting to operate the 

VLS 66xx Edge. 

I. Device Assembly and Disassembly 

1. Select the appropriate VLS video blade or video wand for the 

patient.  
2. Position the blade or wand connection point to the video handle 

mounting block and snap into position by pushing down on the tip of 

the blade or wand.  
3. If using video wand, slide disposable sheath over wand until it 

clicks into place. To remove sheath, unhook from locking tabs and 

slide off.  
4. Removal of the video blade and wand from handles is 

accomplished through reverse of action in Step 2: pulling up on the 

tip of the blade and separating the blade connection point from the 

handle mounting block.  
5. VLS 6630 EDGE PLUS (ONLY) WITH REMOVABLE MONITOR: 

Attach monitor to handle by aligning the arrow on the monitor stem 

with the blue mark on the handle connector. Gently push- do not 

force- the stem into the connector. The stem will lock into position 

when fully inserted. Disassembly involves pulling back on the 

movable sleeve of the stem and gently separating the monitor from 

the handle.  

II. Operating the VLS 66xx Edge Monitor  

Startup 

1.  Press, briefly hold and then release the PWR key in the lower left 

hand corner of the monitor. The system will then power up and 

display the IntuBrite logo for a few seconds, and will then be fully 

functional. If a video blade is appropriately fitted to the handle, 

an image from the camera will appear on the screen. 

2.  Upon turning on the monitor, for a few seconds the screen will 

project in the lower left hand corner a small battery symbol icon 

Summary of Contents for INTUBRITE VLS 6600 Series

Page 1: ...1 IntuBrite VLS 6600 Series Edge Video Laryngoscope Operating Instructions Salter Labs 30 Spur Drive El Paso TX 79906 USA www salterlabs com 800 421 0024 customercare salterlabs com L093 0215 Rev E...

Page 2: ...ssembly 3 II Operating the VLS 6600 Series Edge Monitor 3 III VLS 6600 Series Edge System Operation 4 IV Cleaning Instructions 7 V Batteries and Battery Maintenance 9 VI Troubleshooting 10 VII Warrant...

Page 3: ...2 pulling up on the tip of the blade and separating the blade connection point from the handle mounting block 5 VLS 6630 EDGE PLUS ONLY WITH REMOVABLE MONITOR Attach monitor to handle by aligning the...

Page 4: ...handle should be appropriately cleaned and or sterilized before use see cleaning instructions in PART IV 2 Anti Fog The sapphire lens of the VLS video blade features an electrical lens heating element...

Page 5: ...mode to delete files or reformat the card manually See Below Accessing images and videos Both still images and video loops can be accessed by pressing the Mode button on the display and advancing thro...

Page 6: ...Note that any duplicate files with that number on the SD Card will be automatically replaced with the new file Auto Record Setup To enable Auto Record feature hold down the PWR and OK buttons together...

Page 7: ...l disinfection or sterilization between uses Laryngoscope video handles and wands should be cleaned with low level disinfectant between patients high level disinfection is optional Always follow hospi...

Page 8: ...e exposed to extreme temperatures during cleaning DO NOT AUTOCLAVE THE VIDEO HANDLE BLADE OR WAND Do not use abrasive cleaning techniques or ultrasonically clean Ensure the handle blade and wand are d...

Page 9: ...terilization equipment manufacturer Table 1 V Batteries and Battery Maintenance The VLS 66xx Edge system is equipped with rechargeable lithium ion batteries These batteries are designed to independent...

Page 10: ...of the batteries should be avoided Battery Charging through the Monitor The VLS 66xx Edge battery may be more conveniently recharged through the monitor charger This is accomplished by plugging the U...

Page 11: ...nated Verify battery is charged and installed correctly with positive terminal inserted first down in handle 3 Image appears on screen but is grainy or dim Ensure that handle battery is adequately cha...

Page 12: ...tive will need the part number s and serial number s of the items for which repair is being requested under the warranty The Customer Care Representative will review the situation and if appropriate i...

Page 13: ...the original VLS system an extended warranty package can be purchased separately and independently for each item under the above terms If a VLS system fails for any reason during the extended warranty...

Reviews: