2.2 Setting Up the Instrument
Set the control temperature for each part of the instrument, auto operation
and current time.
Perform these settings on the control panel display of the Embedding Module.
2.2.1 Temperature Setting for Each Part of the Instrument
The heating and cooling parts of the instrument are temperature-controlled
at their respective set values. Refer to
Setting the Temperature
in this
paragraph for information on how to change the set temperatures.
NOTE:
Shown above is an example of the Auto screen.
Temperature setting
location
Setting range
Temperature
display range
Factory setting of
the instrument
Remarks
[1] Paraffin chamber
50 to 75°C
-99 to 99°C
62°C
[2] Left warming chamber
50 to 75°C
-99 to 99°C
62°C
[3] Right warming chamber
50 to 75°C
-99 to 99ºC
62°C
[4] Hot plate
50 to 75°C
-99 to 99°C
62°C
A temperature is displayed
separately for the left and right, but
the set temperatures are the same.
[5] Cooling plate
-10 to 0°C
-99 to 99°C
-5°C
[6] Cold spot
Cannot be set.
Not displayed.
NOTES:
• The forceps holder temperature is set to 8˚C above the set
temperature of the hot plate. Note, however, the upper limit is
80°C.
• At a room temperature of 25°C, no wind and under no load
(nothing on the cold spot), the cold spot temperature should be
between 0 and 15°C when each temperature is set to 62°C.
[6]
Summary of Contents for M01-021E-01
Page 5: ...Table of contents 5 ...
Page 6: ...1 Introduction ...
Page 18: ...2 Installation ...
Page 32: ...3 Explanation of Operations ...
Page 34: ...4 Error Log ...
Page 36: ...5 Troubleshooting ...
Page 46: ...6 Replacement Parts ...
Page 48: ...7 Glossary of Terms ...
Page 52: ...52 ...
Page 53: ...M01 022E 01 ...
Page 54: ...Please visit our websites www sakura finetek com www sakuraus com www sakura eu ...