4
3.2.2TURN OFF
PRESS [OFF] KEY.
3.2.3AUTO-OFF
WHEN POWER AUTO OFF IS SET,AFTER 10 MINUTES IN ZEROING, AUTO-OFF.
(
SEE 3.9.3
)
3.2.4LOW VOLTAGE AUTO-OFF
WHEN THE VOLTAGE LOWER THAN 5.6V, THEN POWER-OFF.
3.3
ZERO
WHEN WITHOUT TARING THE WEIGHT DISPLAY I
S ≤3% MAX. CAPA., PRESS THE
[ON/ZERO] KEY, THE D
ISPLAY SHOWS “ 0 ”, THE ZERO LAMP LIGHTING.
3.4WEIGHING
PUT WEIGHT TO PAN.
3.5TARE WEIGHING
PLACE ONE CONTAINER ON PLATE, PRESS THE [TARE] KEY, THE WEIGHT DISPLAY
IS “ 0 ”,
THE “ZERO” LAMP OFF, THE “TARE” LAMP LIGHTING. PICK UP THE CONTAINER, THE
WEIGHT DISPLAY IS NE
GATIVE, THE “ZERO” LAMP LIGHTING, PRESS THE [TARE] KEY, THE
“TARE” LAMP OFF.
3.6SAVE POWER FUNCTION
ZERO AFTER 40 SECONDS, AUTO SAVE POWER,
SHOWS “ 0 ”. PUT A WEIGHT START
WEIGHING
3.7BATTERY LAMP
“HIGH” MEANS VOLTAGE HIGHER THAN 6.3V, “MID” MEANS VOLTAGE BETWEEN 6V
AND 6.3V, “LOW” MEANS VOLTAGE LOWER THAN 6V.
3.8CHARGE
POWER SUPPLIED BY A INTERNAL RECHARGEABLE LEAD ACID BATTERY (6V/2.3AH).
WHEN IT SHOWS “-BATLO-”. PLEASE TURN OFF AND CHARGE IMMEDIATELY. THE TIME OF
FULL CHARGING IS 12 HOURS.
(
NEED NOT TURN ON WHEN CHARGING
)
3.9
SET AND USE PARAMETER
PRESS <SET/0~9>3 SECONDS
,
ENTER SET AND SHOW MAIN MENU
,
PRESS <SET/0~9> CONTINUOUS
,
IT DISPLAYS
:
RANGE(SET LIMIT)
UNITS(CHOOSE UNIT)
A-OFF
(
CHOOSE AUTO POWER OFF
)
FILT(SET DISPLAY MODE)
ZERO(SET AUTO ZERO)
BUZZLE(SET BUZZLE)
D(CHOOSE DIVISION)