7
4.1.2
PLEASE CALIBRATE SCALE REGULARLY.
4.1.3
DO NOT OVERLOADING. DO NOT THROW HEAVY THINGS ON THE SCALE VIOLENTLY. THE
DAMAGE FOR OVERLOADING IS NOT WITHIN THE RANGE OF GUARANTEED
MAINTENANCE.
4.1.4
DON’T TOUCH CHEMICALS SUCH LIKE SOLVENT, ETC.
4.1.5
KEEP THE SCALE CLEAN IN ORDER TO WEIGHING WELL.
4.1.6
THERE IS A LEAD SEAL ON THE BOTTOM OF THE SCALE. USERS ARE NOT ALLOWED TO
REMOVE THE SEAL. IF THE SEAL IS DAMAGED THE SCALE IS NOT GUARANTEED FOR
MAINTENANCE.
4
.2
FOR SEALED LEAD ACID BATTERY
4.2.1THE WORKING TEMPERATURE OF BATTERY IS -10
℃
~+40
℃
. IT CAN BE CIRCLE CHARGED
ABOUT 300 TIMES AT 20
℃
.
4.2.2 PLEASE CORRECTLY CONNECT LINE BETWEEN ADAPTER AND BATTERY, BLACK TO BLACK,
RED TO RED.
4.2.3PLEASE CHARGE IN TIME. IF YOU DON
’
T USE IT IN A LONG TIME, PLEASE
CHARGE AT LEAST EVERY 3 MONTHS.
4.2.4
IF THE USING TIME OF BATTEY BECOMES SHORT, PLEASE CHANGE NEW ONE.
4.2.5 IT HAS A ONE MONTH GUAANTEE.
5
ADJUSTMENT
1.
PUSH THE ADJUST KEY TO SWITCH ON, THE SWITCH IS ON THE BOTTOM OF
THE SCALE, LOCKED WITH A SEAL .PRESS
【
ON/ZERO
】
KEY, AFTER A WHILE
START TO ADJUST
;
IF SHOW
ERR01,ADJUST SWITCH IS OFF
2.
PRESS
【
SET/0~9
】
AND
【
ON/ZERO
】
KEY THE SAME TIME, CHOOSE WEIGHING
UNITS-KG, LB OR COUNTING FUNCTION, PRESS
【
TARE
】
TO CONFIRM. IF
DON’T CHOOSE LB, PRESS
【
SET/0~9
】
, IT DISPLAYS PCSH9, PRESS
【
TARE
】
TO CONFIRM. THEN THE
DISPLAY SHOWS “C”. “C” IS THE MAX. CAPACITY,
PRESS
【
SET/0~9
】
TO CHOOSE C-300,C-600,C-1500,C-3000,C-6000,C-15000,
AND PRESS
【
TARE
】
TO CONFIRM
;
3.
NOW IT DISPLAYS 1P(ONE POINT CALIBRATION) OR 3P(THREE POINTS
CALIBRATION)
,
PRESS SET/(0~9)
】
TO CHOOSE 1P OR 3P
,
PRESS
【
TARE
】