Acoustic Fluid Logger and Pressure Pulse Gas Gun
13
6. Fire the Pressure Pulse Gun by sliding the shuttle valve to the fire position. This is easily
accomplished by pressing the shuttle valve with the heel of the hand.
7. The paper tape begins to print an echo trace after the gun is fired. Let the tape run until a fluid
kick is observed. Collars inside the well will register on the tape, but the deepest kick will be the fluid
level. After the 4 - C - Start (1/4 compressed) printout determines a fluid level, press “Stop” to stop the
printout, then press “Print” to print a full-length tape.
Note: The Acoustic Fluid Logger III stores seven representations of each echo internally,
each time the gun is fired. These echoes are ready to be printed individually, each offering different
amplification levels of the well data. This feature takes into account the fact that all wells are different,
with different variables affecting the operator’s ability to see the fluid level data. With differing
amplification, the operator has seven different options that can be printed, and seven ways to look at the
same fluid level. (See the Keypad and Interface section of this manual.)