USER MANUAL
EvaluationTools
STIM300 STIM377H Evaluation Kit
DOK453 rev.2
Page 26 of 29
7.7
Logging panel
Panel content
Functionality and description
Start button
Starts data logging
Stop button
Stops data logging
Stop criteria slide
User can
select between “Manually”, “No of samples” and “Time
elapsed” for stopping a measurement series
Samples box
Used for defining number of samples when logging a finite number of
samples
Average
Used for downsampling of data. Average value of selected number of
values is logged to file
Time elapsed
Shows the time elapsed since start of test
Samples acquired
Shows number of samples acquired
CRC_errors
Shows number of CRC errors (normally 0, otherwise the user should
consider to reject results data in any analysis)
Resynch’s
Increments from 0 to a number if any re-synchronisations are needed
in order to re-establish data collections from module
Table 9: Logging panel descriptions
Log to file capability:
Quad core processor is recommended when measuring on two gyro modules simultaneously
The size of the log file is only limited by the available space on the storage media in use
The path for result file storage is defined in the active parameter file
The program should be run with administrator rights to ensure the creation and storage of the result file
7.8
Parameters panel
Panel content
Functionality and description
===== General parameters =====
Password
Current valid password to be able to edit the parameters list. The
default
password is “stim”
Folder for result-file storage
Path to storage (e.g.
“c:\userdata\test\")
What priority will this program run with
Instructs the program priority for the PC operation system
What format to use for result files
ASCII text by default. Can be changed to 8 byte binary
Name of file with language definitions
Application can be configured with language other than English
===== Device communication =====
IMPORTANT MESSAGE: Always verify
hardware connections and COM port settings
before trying to connect to the device
RS422 port # to device 1
Defining which COM port # to assigned to gyro module # 1
RS422 port # to device 2
Defining which COM port # to assigned to gyro module # 2
RS422 Bitrate [bit/s]
RS422 bit rate selection
RS422 Stopbit
1 or 2. Default is “1”
RS422 parity
None, odd or even. Default is “None”
===== External Hardware =====
The GPIB-card # to use
Interface for external power supply (optional). If card(s) are in use;
the first card will be assigned to #0, second to #1, etc. Default value
is “0”
Type of power supply used
External power supply (optional). Default "None" (not in use).
Agilent E3631A, E3633A and E3644A are supported
Interface that the power is connected with
Interface type for external power supply (optional). Default "None"
(not in use). RS232 (for Agilent E3631A only) and GPIB supported
Port or address to power
GPIB port for external power supply (optional). Default "0" (not in
use). Selectable up to 31