37
use of a mask and glasses to apply the bicarbonate jet.
1. Pressing the (01) key a second time the Bicarbonate Jet function (B) will be activated. Proceed
according to the instructions below.
2. Remove the top cap (G) by unscrewing it (according to the drawing) and add enough sodium
bicarbonate for a section of prophylaxis, that is, from 20 to 40g (do not exceed the level indicated
on the container). The bicarbonate level is visible through the transparent container (H). To remove
the leftover bicarbonate powder, unscrew the container (H) (as shown) and clean.
The volume of water and air flow can be regulated according to the need, as follows:
3. Direct the bicarbonate jet handpiece to a container (Ex: spit pan, sink bowl, etc.).
4. Operate the foot control and adjust the water volume “water selector (D)” and air flow “air
selector (E)”. The amount of water in excess will cause a decrease in the effect of the powder due
to washing. Decreasing the water too much will cause the powder to become more aggressive.
01
E
G
H
B
Do not add more than 40g of bicarbonate to the container (H) as this will cause
the powder to become blocked.
The effectiveness depends on the dosage of the water volume and the amount
of powder.
Summary of Contents for Sonic Duo
Page 1: ......
Page 2: ...2 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Accessories that do not accompany the product 01 07 13 04 10 02 08 14 03 09 15 05 11 06 12 ...
Page 56: ...56 10 6 EQUIPMENT DIMENSIONS 10 6 1 Sonic Duo ...
Page 57: ...57 10 6 2 Sonic Duo Fit ...
Page 58: ...58 10 6 3 Sonic ...
Page 59: ...59 10 6 4 Sonic Fit ...
Page 65: ...65 ...
Page 66: ...66 ...
Page 67: ...67 ...
Page 68: ...68 ...