background image

10

where P is the maximum output 
power rating of the transmitter in 
watts (W) according to the 
transmitter manufacturer and d is the
recommended separation distance in 
metres (m). Field strengths from 
fixed RF transmitters, as determined 
by an electromagnetic site survey,a 
should be less than the compliance 
level in each frequency range.b 
Interference may occur 
in the vicinity of 
equipment marked 
with this symbol:

NOTE 1 

At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 

These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and l
and mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted 
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the location in which the
“Daylight Lamp” is used exceeds the applicable RF compliance level above, the Medical Daylight should be 
observed to verify normal operation. If abnormal performance is observed, additional measures may be 
necessary, such as reorienting or relocating the “Daylight Lamp”.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [

V

1

] V/m.

Immunity test

IEC 60601

Compliance

Electromagnetic environment –

test level

level

guidance

RF emissions

Group 1

The “SAD Solutions Bluelight” uses RF energy only for it internal 

CISPR 11

function. Therefore, its RF emissions are very low and are not likely 
to cause any interference in nearby electronic equipment.

RF emissions

Class B

CISPR 11

Harmonic
emissions

Class A

IEC 61000-3-2

Voltage
fluctuations/

Complies

flicker emissions
IEC 61000-3-3

Emissions test

Compliance

Electromagnetic environment – guidance

The “SAD Solutions Bluelight” is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.

emc information

Guidance and manufacturer’s declaration – electromagnetic immunity

The “SAD Solutions Bluelight” is intended for use in the electromagnetic environment specified
below. The customer or the user of the “SAD Solutions Bluelight” should ensure that it is used in such
an environment.

SAD Solutions IB  16/9/10  12:31  Page 10

Summary of Contents for BLUElight

Page 1: ...BLUEl ight LED Portable Light Therapy The natural way to improve energy levels and enjoy that feel good factor instruction guide SAD Solutions IB 16 9 10 12 31 Page 1...

Page 2: ...uses the latest blue light technology and is a leap forward in the treatment of Seasonal Affective Disorders and other Circadian Rhythm related problems such as Jet Lag sleep disorders and problems b...

Page 3: ...nsity than full spectrum light and therefore much kinder on the eyes SAD Solutions are a UK company that use only the very best manufacturing techniques the highest quality components and the best des...

Page 4: ...ck plug Do not pull the cord I Always hold the adapter when unplugging it from a socket Do not pull the cord I You should not use the adapter if there are any signs of damage to the cord symbol indica...

Page 5: ...t session timer indicator Battery capacity charge indicator LED intensity indicator Sun Symbol will always remain on No bars indicates 20 level Product should be charged No bars flashing indicates bat...

Page 6: ...The environment where you are using BLUElight should be a well lit room Natural daylight or electric lighting if it is dark Q You can carry on doing normal activities whilst using BLUElight ie Reading...

Page 7: ...e terms this means the body wakes up and all functions are fully on during daylight hours During darkness the body slows down and all functions reduce until it goes into sleeping mode Q This is why yo...

Page 8: ...at your destination in the morning rather than the evening Sleep on the plane if you are able Then try to keep awake for as long as possible after you reach your destination preferably in natural dayl...

Page 9: ...IEC 61000 4 11 60 dip in UT 60 dip in UT mains interruptions it is for 5 cycles for 5 cycles recommended that the SAD Solutions Bluelight be powered from 70 UT 70 UT an uninterruptible power supply o...

Page 10: ...Daylight should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the Daylight Lamp Over the frequency...

Page 11: ...e SAD Solutions Bluelight The SAD Solutions Bluelight is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the SAD Solutio...

Page 12: ...ccident or misuse this is not included in our guarantee Important This undertaking is in addition to consumer s statutory rights and does not affect those rights in any way repairs and services If the...