background image

The pressure measuring transducer 

PREMASGARD

®

 SHD-LC

 measures relative pressures in the bar range.  

It converts the measurand pressure into standard signals of 0  -10  V or 4...20 mA. Process connection is G ½ " straight external pipe thread.  
SHD-LC is used for pressure measurement in gaseous and liquid media. Applications of this pressure transmitter are in hydraulics,  
pneumatics, process technology, in mechanical and plant engineering. 

Not applicable for ammonia and Freon!

TECHNICAL DATA:

Power supply: ........................ 24 V  AC ⁄ DC for output 0 -10 V   

7 - 33 V  DC  for  output  4 ... 20 mA

Measuring ranges: .................. see table 

(other ranges upon request)

Output signal: ........................ 0 -10 V, 3-wire, (working resistance > 10 kOhm)  or 

4... 20 mA,  2-wire,  (working  resistance  <  (UB  (V) -7 V)  ⁄  0,02 A ;   
R

 depending on working resistance

Electrical connection: .............. plug connector DIN EN 175301-803-A
Pressure connection: ............... G ½ " sealing at the back,  

and manometer (combined) with profile gasket FPM,  
special WW G ¼ "  DIN  3852

Type of pressure: .................... relative
Measuring principle: ................

ceramic measuring cell 

Temperature of medium: .......... –15...+ 125 °C
Mounting: .............................. directly on pressure line
Enclosure: ............................. stainless steel 1.4305
Connecting head: .................... plastic,  approx.  98 x 50 x 34 mm
Medium contacting parts:......... stainless steel type 1.4305, AISI  303; 

measuring element ceramic Al

2

0

3

 (96%); 

sealing material FPM (Viton)

Response time: ...................... 2 ms   (1 ms  typical)
Load changes: ........................ <  100 Hz
Characteristic line: ................. ± 0.3 %
Overload range/
Bursting pressure: .................. <  4 bar:   3 x  FS 

>  4 bar:   2.5 x  FS

Protection class: .................... III  (according to EN 60 730)
Protection type: ..................... IP 65  (according to EN 60 529)
Standards: ............................ CE conformity, electromagnetic compatibility  

according to  EN 61 326, EMC directive 2004 ⁄ 108 ⁄ EC

ACCESSORIES: ....................... incl. receptacle DIN EN 175301-803-A

PREMASGARD

®

    SHD - LC 

Rev.  2015 - 1.0  GB

Connecting diagram 

SHD - LC - U

Connecting diagram 

SHD - LC - I

1

2

3

IN

+

OUT

GND

3

2

1

Output pressure 0-10V
GND
Supply voltage

UB+ 24V AC/DC

3

2

1

1

2

3

IN

+

OUT

Output pressure 4...20mA
Free
Supply voltage

UB+ 24V DC

Type ⁄ WG1

Measuring  

Range

Item No.

SHD -LC - U

U - variant

SHD -LC -U 6

0 … 6  bar

1301-2121-0550-120

SHD -LC -U 10

0 …10  bar

1301-2121-0560-120

SHD -LC -U 16

0 …16  bar

1301-2121-0570-120

Type ⁄ WG1

Measuring  

Range

Item No.

SHD -LC - I

I - variant

SHD -LC -I 6

0 … 6  bar

1301-2122-0550-000

SHD -LC -I 10

0 …10  bar

1301-2122-0560-000

SHD -LC -I 16

0 …16  bar

1301-2122-0570-000

Summary of Contents for PREMASGARD SHD

Page 1: ...ageanleitung Druckmessumformer incl DIN Steckdose mit aktivem Ausgang G Operating Instructions Mounting Installation Pressure measuring transducers incl DIN receptacle with active output F Notice d in...

Page 2: ...PREMASGARD SHD PREMASGARD SHD LC D G F r Ma zeichnung SHD Dimensional drawing SHD LC Plan cot G1 2 PG11 SW24 85 115 20 3 30 50...

Page 3: ...ber hrende Teile Edelstahl Stahltyp 1 4305 Ansprechzeit 2 ms typ 1 ms Kennlinie 0 3 berlast 6 bar 5 x EW 6 bar 3 x EW max 1500 bar Berstdruck 6 bar 10 x EW 6 bar 6 x EW max 2500 bar Schutzklasse III n...

Page 4: ...ung Geh use Edelstahl Stahltyp 1 4305 Anschlusskopf Kunststoff ca 98 x 50 x 34 mm messstoffber hrende Teile Druckanschluss aus Edelstahl Stahltyp 1 4305 AISI 303 Messelement aus Keramik Al203 96 Dicht...

Page 5: ...ausgeschlossen Die Installation der Ger te darf nur durch Fachpersonal erfolgen Es gelten ausschlie lich die technischen Daten und Anschlussbedingungen der zum Ger t gelieferten Montage und Bedienung...

Page 6: ...34 mm Medium contacting parts stainless steel 1 4305 Response time 2 ms 1 ms typical Characteristic line 0 3 Overload range 6 bar 5 x of final value 6 bar 3 x of final value max 1500 bar Bursting pres...

Page 7: ...125 C Mounting directly on pressure line Enclosure stainless steel 1 4305 Connecting head plastic approx 98 x 50 x 34 mm Medium contacting parts stainless steel type 1 4305 AISI 303 measuring element...

Page 8: ...ed from warranty or liability These devices must be installed by authorised specialists only The technical data and connecting conditions of the mounting and operating instructions delivered together...

Page 9: ...8 x 50 x 34 mm Parties en contact avec le milieu acier inox de type 1 4305 Temps de r ponse 2 ms temps caract ristique 1 ms Ligne caract ristique 0 3 Surcharge 6 bars 5 x pleine chelle 6 bars 3 x plei...

Page 10: ...us pression Bo tier acier inox de type 1 4305 T te de raccordement mati re plastique env 98 x 50 x 34 mm Parties en contact avec le milieu raccord de pression en acier inox de type 1 4305 AISI 303 El...

Page 11: ...et appareil L installation des appareils doit tre effectu e uniquement par un sp cialiste qualifi Seules les donn es techniques et les conditions de raccordement indiqu es sur la notice d instruction...

Page 12: ...e UB 24V AC DC 3 2 1 1 2 3 IN OUT Output pressure 4 20mA Free Supply voltage UB 24V DC WG1 01 SHD U U SHD U 1 0 1 1301 2111 0520 220 SHD U 2 5 0 2 5 1301 2111 0530 220 SHD U 6 0 6 1301 2111 0550 220 S...

Page 13: ...EC DIN EN 175301 803 A r PREMASGARD SHD LC Rev 2015 1 0 RU SHD LC U SHD LC I 1 2 3 IN OUT GND 3 2 1 Output pressure 0 10V GND Supply voltage UB 24V AC DC 3 2 1 1 2 3 IN OUT Output pressure 4 20mA Free...

Page 14: ...r ZVEI VDE EMV...

Page 15: ...ions techniques Copyright by S S Regeltechnik GmbH Nachdruck auch auszugsweise nur mit Genehmigung von S S Regeltechnik GmbH gestattet Reprints in part or in total are only permitted with the approval...

Page 16: ...PREMASGARD SHD PREMASGARD SHD LC D G F r...