700-1000 - B
June - 2010
15
MAINTENANCE and CLEANSING
Minimal maintenance will prolong the life of your lift. Keep all electronic components clean and dry. Keep
the Console/Battery Cover installed at all times to prevent moisture from collecting on the Control Box
and battery. Excessive moisture collection can affect battery and lift performance and could lead to
battery failure and/or the lift failing to operate. If the lift is used outdoors, an optional full cover is available
and recommended.
Owners of lifts shall be aware of any applicable local, state/provincial or federal/national regulations
regarding the inspection and or testing of lifts.
The following schedule must be performed to insure proper operation with the Daily items
performed before each use:
Maintenance Performed
Daily
Weekly
Monthly
Check battery level before each use / Charge battery daily
Wipe Control Box and battery connection with a clean dry rag
Examine lift for any damage, loose or missing hardware
Test for normal operation
Spray gear assembly with a heavy-duty rust inhibitor/lubricant
such as LPS 3 - Heavy-Duty Inhibitor
Make sure all cable connections are properly secured
Inspect lift frame, mast, support arm and seat assembly for rust
Inspect weight stack for rust/treat with touch up paint
Cleansing Performed – after each use
Rinse seat and seatbelt/stability vest with fresh water between
each use - Cleanse seat and seatbelt/stability vest with a
disinfectant solution of 1:100 dilution of household bleach to
fresh water and then rinse with fresh water and dry entire lift
daily. In the event of a contamination incident such patient/user
excreta - cleanse seat and seatbelt/stability vest immediately
with the disinfectant solution*
Cleanse all battery connections with a nylon scouring pad
Cleanse all metallic surfaces with a cleaner wax to maintain the
finish of the lift
*
When using the disinfection solution avoid direct contact with the skin and eyes. In the event of a
contamination incident - immerse the seat belt or stability vest in the disinfection solution for 10 min. and
then rinse thoroughly with fresh water.
TROUBLE SHOOTING
Be sure the battery is fully charged before troubleshooting.
Lift does not rotate
Does lift raise or lower?
Yes.
1. Check connection to Control Box. Be sure plug is pushed in all the way.
2. Check hand control connection to Control Box for damaged pins.
3. Check connections on terminal block located on frame for loose wires.