START-UP
Axio Lab.A1
Installing standard components
Carl Zeiss
04/2013 430037-7144-001
51
3.1.6
Fitting and removing push&click modules in the reflector turret
The reflector turret 4x stand transmitted and reflected light (BioMed) and the reflected light stand
(Material) is firmly installed.
The modules must be inserted and removed from the front after removing the cover cap.
Inserting a module:
x
Remove cover cap (Fig. 3-9/
4
) from the stand
towards the front.
x
Insert the module (Fig.
3-9/
2
) as illustrated
together with the retaining brackets on the
right and left (Fig. 3-9/
3
) diagonally from below
into the upper spring clips (Fig. 3-9/
1
) of the
reflector turret.
x
Then apply pressure to the module from below
until it also securely engages with the lower
spring clips of the reflector turret.
The position number of the P&C module is
shown on the right hand side of the reflector
turret adjacent to the position of the respective
P&C module.
x
Apply the supplied stickers with the filter
combination data of the respective module to
the corresponding field of the cover cap
(Fig. 3-9/
5
, position 1 to 4).
Removing a module:
x
Slightly tilt the module in order to detach it
from the lower spring clips, then from the
upper spring clips of the reflector turret.
x
Once the reflector modules have been installed /
removed, refit the cover cap. The cover cap
should be fitted onto the stand as straight as
possible to avoid the knurled ring of the
reflector turret becoming jammed and
damaged.
x
Apply pressure to the cover cap until the
retaining brackets have engaged.
Fig. 3-9
Replacing the reflector module