Applicable to cats, dogs, monkeys, pigs and other animals of 100kg weight or less;
Two options available: close and semi-open anesthesia mode;
The oxygen flow meter adjustable ranges from 0.1 to 4 L/min;
With oxygen flush function to replace the system with pure oxygen at a speed of
10L/min;
CO
2
absorbent canister (2100 ml) is installed in the front part of the machine that is
easy to observe and dissemble for changing;
Newly designed vaporizer provides enough flow of anesthetic gas for animal.
Support both Cage Mount and Selectatec.
The concentration of anesthetic gas from the vaporizer can be adjusted from 0 to
5 % (Isoflurane) and 0 to 8% (Sevoflurane), without effect by fluctuation of flux,
temperature, flow rate and pressure. The safety locking device well prevents
accidental volatilization of anesthetic agent;
Compact and easy to clean.
1.5 Facility requirements
Table 1-1 provides the facility requirements necessary to ensure reliable operation and safety of
the anesthesia system.
Table 1-1 Facility Requirements
Specification
Work Environment
Temperature:10
℃
~35
℃
Humidity:15%~95% (
non-condensing
)
Air pressure:57kPa~106kPa
Storage Environment
Temperature:-20
℃
~60
℃
Humidity:10%~95% (
non-condensing
)
Air pressure:50kPa~106kPa