27
8-Product information
8.1 Product warranty
This warranty is only applicable to those new products purchased from RWD or dealers authorized
by RWD, as well as the first person to whom it extends.
8.2 Product specifications
Material
Mainly aluminum alloy
Working condition
Temperature
:
10 ~ 35
℃
Humidity
:
15 ~95 %
(non-condensing)
Storage condition
Temperature
:
-20 ~ 60
℃
Humidity
:
10 ~ 95 %
(non-condensing)
Oxygen flow rate
0.2~10 L/min, 0.2~8 L/min when concentration >4 %
Oxygen grade
Medical grade
Concentration range
Isoflurane
:
0.5
~
5 %(v/v), Sevofluran: 0.5
~
8 %(v/v)
Settable concentration
point
Isoflurane: 0 ~ 0.5 ~ 1.0 ~ 1.5 ~ 2.0 ~ 2.5 ~ 3.0 ~ 3.5 ~ 4.0 ~
4.5 ~ 5.0 %(v/v)
Sevofluran: 0 ~ 0.5 ~ 1.0 ~ 2.0 ~ 3.0 ~ 4.0 ~ 5.0 ~ 6.0 ~ 7.0 ~
8.0%(v/v)
Perfusion volume of
anesthetic agent
The recommend volume is 100 mL, as the volume between
max and min visible liquid level is about 120 mL.
Consumption of
anesthetic agent
About 3 × oxygen flow rate (L/min) × set concentration
value(% (v/v))
e.g. When the isoflurane concentration is set at 2 % and the
oxygen flow rate is set at 600 mL/min, a bottle of isoflurane
of 100 mL could be consumed for about 28 hours.
Loss of anesthetic agent
22
℃,
dial at 0 %
,
< 0.5 mL/24h.
Max pressure load
50 kPa
Max inclined angle
30°