![RWD D180P User Manual Download Page 7](http://html1.mh-extra.com/html/rwd/d180p/d180p_user-manual_1503442007.webp)
× width × height:
Capacity
~180L
Weight
~120kg
Heating method
Inner wall heating, air casing
1.7 Settable parameters
Parameters
Detailed description
Temperature
10.0-50.0°C
CO
2
concentration
0.0-20.0%
Temperature alarm threshold
Upper limit: (T+0.1) - (T+5.0)
Lower limit: (T-0.1) - (T-5.0)
CO
2
concentration alarm threshold
Lower limit: (C-0.1) - (C-5.0)
Upper limit: (C+0.1) - (C+1.0)
Lower limit of humidity alarm
Lower limit: 80-100
%
HEPA filter screen countdown
Adjustable range: 0 - 365 day(s)
1.8 Product list
Attachment
Accessory name
Quantity
Application description
Standard
Host
1
Body of CO
2
Incubator
Standard
Vertical plate of CO
2
Incubator (provided
with 6 button holes)
2
For supporting stainless steel shelves
Standard
Stainless steel shelves
3
Carrier for culture dishes (up to 15
optional shelves, each with load
bearing of 7kg)
Standard
5.5 inch HEPA filter
1
Air filtering
Standard
Humidity pan
1
Holding water to increase humidity
Standard
Components of stacking
4
For stacking and fixing of equipment
Standard
Access port filter
2
For air filtering
Standard
Ejector rob assembly
3
For supporting the vertical plate of
CO
2
Incubator