4.6 Sensor calibration
After the equipment stably working, user could calibrate probes regularly as
needed. The frequency of calibration depends on usage, environment conditions and
accuracy requirements.
Good laboratory specifications require at least one calibration of probes each
year for the equipment.
Click [Cali] on the [Menu] interface to enter the following interface:
External auxiliary calibration of temperature probe
The chamber temperature must be stable before calibration. Place the prepared
temperature calibration instrument in the center of the incubator chamber. The
instrument should be within the airflow, rather than being against the shelf. After
the calibrated temperature value is stabilized, the measured value of the
calibration instrument is input to the [Current Value] of the temperature probe by
the “+/-” button or the numeric keypad.