42
Operating instructions for PODOLOG NOVA 3s | REV. 0
8.2
INFORMATION REGARDING ELECTROMAGNETIC COMPATIBILITY
Guidelines and manufacturer’s declaration – electromagnetic interference
The PODOLOG NOVA 3s is designed for use in electromagnetic environments as specified
below. The customer or client using PODOLOG NOVA 3s hould ensure that the device is
used in the prescribed environment.
Emission measurements
Conformance
Guidelines for the electro-magne-
tic environment
HF-emissions according
to CISPR 11
Group 1
The PODOLOG NOVA 3s
uses HF solely for internal func-
tions. Its HF emissions are very
low and it is unlikely
that any interference will be
caused to nearby electronic
devices.
HF emissions
according to CISPR 11
Class B
The PODOLOG NOVA 3s is suit-
able for usage on all premises,
which are directly connected to
the public power supply.
Harmonic components ac-
cording to IEC 61000-3-2
Class A
Voltage fluctuations / flicker
according to IEC 61000-3-3
Fulfilled
Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper operation.
If such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally."
Use of accessories, transducers and cables other than those specified
or provided by the manufacturer of this equipment could result in increa-
sed electromagnetic emissions or decreased electromagnetic immunity
of this equipment and result in improper operation.