Quick Guide 

 

9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 

 

2(4) 

v.2017.3D 
 

 

 

 
 
 

l.

 

When starting one of the test a set-up instruction (hint) 
will pop-up, see image.  
 

m.

 

 Press the right arrow key to get to the measure screen.  

 

 
 
 
 
 
 
 
 
 

Notice: 

Depending on configuration of Ocean, a Position Check may be 
suggested 

(yellow bar besides the green “ready” box)

. If so, just 

abort the Position Check by clicking anywhere in the yellow 
area

.  

This will be the case for every new target/filter selection. 
 

With the good geometry of the Pristina unit, and use of the 
Piranha Pristina holder, GE part # 5758284, the Position check 
does not add confidence to the measurement. 

 

Conversion Factor 

 CAL 009 

The image to the right shows the template for the conversion 
factor measurement. 
Follow the GE job card instructions for the measurement.  
 

Notice:  

The template is built for mR units. If mGy units shall be used, 
just right click on the unit and select mGy.  
 
When ready for next test, just press the 

New Quick Check 

button. You shall now be back to the selection screen as 
described under (i) above.  
 
 
 
 

HVL 

 CAL A010 

The template is built for the sequence to measure small and large focus in the same sequence, but can be 
used also for one at the time.  
The template can be used for one shot HVL as well as for traditional measurements by adding layers of 
aluminum sheets. 
 
 

 

 
 

 
 
 