background image

 

Quick Guide 

 

9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 

 

3(4) 

v.2017.3D 
 

 

 

 

 
The image to the right shows the template for the HVL 
measurement on Small Focus.  
The template supports one shot HVL as well as traditional method 
by adding layers of Aluminum. 
Follow the job card instructions for the measurement. No settings 
in Ocean should have to be made. 
 

Notice:  

The

 

Position check may be suggested (yellow bar besides the green 

“ready” box)

. Is so, abort by clicking in the yellow field as described 

above. 
 
When ready for next test, just press the 

New Quick Check 

button. 

You shall now be back to the selection screen as described under (i) 
above, or exit Ocean.  
 

Tube Voltage 

 ELE 010 

The template is built for measurement without compression 
paddle. The compression paddle may be used. Then you need to go 

Meter Settings¸ 

and change the setting to 

Use 

compression paddle.  

 
Follow the GE job card instructions for the measurement.  
 

Notice:  

The

 

Position check may be suggested. If so, abort by clicking in the 

anywhere in the yellow area as described above. 
 
When ready for next test, just press the 

New Quick Check 

button. 

You shall now be back to the selection screen as described under (i) 
above, or exit Ocean.  
 
 

3.

 

Measure using GE Mammo Quick Check  

Setup 

n.

 

Place the Black Piranha in the holder as shown in image (right) with the left pin 
inside the tripod mounting hole. 

The Black Piranha will only fit in one way.  

o.

 

Slide the holder into the GE dosimeter support. 

p.

 

The Black Piranha can be left in this positon for all measurements described below. 

q.

 

Follow the job card instructions when it comes to image detector protection, IQST 
phantom set-up, compression paddle set-

up, etc…

 

 

Measure 

When starting Ocean the measure guide as shown in image (right) should appear. If not, the 

option “Auto

-

start Quick Check” is deactivated. This function can be activated under Options. 

 

 

r.

 

Select 

Mammo quick check.

 

s.

 

A set-up instruction (hint) will pop-up. 

t.

 

Press the right arrow key in the lower left corner to get to the measure screen.  

 
 

 

 

 