background image

 

Quick Guide 

 

9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 

 

4(4) 

v.2017.3D 
 

 

 

 

u.

 

Select calibration, by clicking the model/beam 
quality. See image to the right.  
As default compression paddle is in use. If 
compression paddle is not in use, go to Meter 
settings and deselect.  
 

Notice: 

The

 

Position check may be suggested (yellow bar 

besides the green “ready” box)

. If so, abort by 

clicking in anywhere in the yellow area as 
described above. 

 

Notice: 

If doing traditional HVL measurements by adding 
layers of Al sheets, the thickness of the filter 
must be entered in the field for Added filtration under Meter Settings.   
 

Notice: 

When measuring the CF factor using the IQST phantom the displayed kV may not be within 34 ±1 kV. This does not 
affect the accuracy of the dose reading which is the primary target of this measurement.   

 

 

4.

 

Measure using standard Ocean configuration  

 
In case the GE Quick Check Templates are not configured 
with the Ocean in use, the procedure is slightly different 
from what is describe under heading 3 above.  
After entering 

Quick Check 

(see image below), the measure 

guide will look a little different. 
 
Follow the guide sequence and select Mammography - GE Healthcare - Pristina - Rh/Ag IQST Pristina - Compression Paddle not 
used. 
 
The image to the right shows the template for the 
conversion factor measurement.  
Follow the GE job card instructions for the 
measurement.  
 

Notice: 

The

 

Position check may be suggested (yellow bar 

besides the green “ready” box). If so, abort by clicking in 

anywhere in the yellow area as described above. 
 

Notice:  

Depending on Ocean configuration the unit will be mR 
or mGy. Unit can be changed - Just right click on the unit 
and select unit.  

 
Notice: 

When measuring the CF factor using the IQST phantom the displayed kV may not be within 34 ±1 kV. This does not affect the 
accuracy of the dose reading which is the primary target of this measurement. 
 
When ready for next test, just press the 

New Quick Check 

button. You shall now be back to the selection guide described above.   

 
For deeper understanding and guidance

, please see User’s Manuals for Piranha and Ocean.

 

 

*** END *** 