Temperature Calibrator (RTD + TC) / EN
_________________________________________________________________________________
03/27/2020 Version No. 1 28
4.7 Auto-Power-Off
Step 1: press SETUP to enter SETUP
Step 2: press
▲ ▼
to select item of AutoPowerOff
Step 3: press ◄►to enable or disable
Step 4: press SETUP again to exit
4.8 Restore Factory Settings
Step 1: press SETUP to enter SETUP
Step 2: press
▲ ▼
to select item of TC 0% or 100%
Step 3: press SHIFT to clear input field
Step 4: press numerical key pad to enter numbers. If value of 1234 is
entered, all the values of CJC (S), CJC (M), RTD 0%, RTD 100%, TC 0%,
and 100% will be restored to factory settings.
Step 5: press SETUP again to exit