![RS PRO RS-136 Instruction Manual Download Page 35](http://html1.mh-extra.com/html/rs-pro/rs-136/rs-136_instruction-manual_1495969035.webp)
Temperature Calibrator (RTD + TC) / EN
_________________________________________________________________________________
03/27/2020 Version No. 1 32
RTD source in
°
C:
Accuracy is based upon 4W connection, driving voltage is less than 1.7V and the
excitation current is based upon 0.5mA to 5mA (0 to 400
Ω
) and 0.05mA to 5mA
(400
Ω
to 7000
Ω
). For 3-wire RTD source, assuming all three RTD leads are
matched, add 1.0 °C (Pt10 and Cu10), 0.6 °C (Pt50 and Cu50), 0.4 °C (Other
RTD types) to the specifications.
RTD Type
(α)
Source (°C)
Range
Accuracy
10
Ω
Pt(385)
-200 to 100
1.5
100 to 800
1.8
50Ω
Pt(385)
-200 to 100
0.4
100 to 800
0.5
100Ω
Pt(385)
-200 to 100
0.2
100 to 800
0.015%+0.18
200Ω
Pt(385)
-200 to 100
0.2
100 to 630
0.015%+0.18
500Ω
Pt(385)
-200 to 100
0.3
100 to 630
0.015%+0.28
1000Ω
Pt(385)
-200 to 100
0.2
100 to 630
0.015%+0.18
100Ω
Pt(3902)
-200 to 100
0.2
100 to 500
0.015%+0.18
100Ω
Pt(3916)
-200 to 100
0.2
100 to 630
0.015%+0.18
100Ω
Pt(3926)
-200 to 100
0.2
100 to 630
0.015%+0.18
10Ω Cu(427)
-100 to 260
1.5
120Ω Ni(672)
-80 to 260
0.15
50Ω Cu(427)
-180 to 200
0.4
100Ω Cu(427)
-180 to 200
0.2
YSI400
15 to 50
0.2
Temperature coefficient
:
(± 0.002% of reading ±0.002% of range)/°C (<18°C or
>28°C)