The forth step of checking leves. “RUN” will appear on LCD and
pressures are checked at 0/30 mmHg. If the pressure difference at 0/30
mmHg is within ± 3mmHg, the calibrator shows "PASS". If "FAIL"
appears, the faulty pressure points will be shown (both the default
pressure check point and the actual pressure measured will appear
alternatively.).
If "FAIL" happens, press the "Auto calibration" key for re-calibration.
After this process is complete, press the "Readout" key for double
check.
check point on
Calibrator
check point from Blood
Pressure Monitor
If deviation under ±3mmHg
If deviation over ±3mmHg
Readout Check is OK
Calibrator Screen
Blood Pressure
Monitor Screen
Calibrator Screen
Blood Pressure
Monitor Screen
Press Auto Calibration Mode
(follow instructions on Page 18)
Calibrator Screen
Blood Pressure
Monitor Screen
Checking the pressures between calibrator and
monitor. If the deviation of Third Pressure Points
is under ±3mmHg, the calibrator would
automatic check Forth Pressure Points
(0/30 mmHg)
Third
Pressure
Points
If deviation between calibrator and
monitor is under ±3mmHg
Forth
Pressure
Points