To check the second pressure levels. “RUN” will appear on LCD and
pressure is checked at 0/210/180/150 mmHg points. If the pressure
difference at 0/210/180/150 mmHg is within ± 3mmHg, the calibrator
will rapid exhaust to 0 mmHg, then go on checking 3rd points. if "FAIL"
appears, the faulty pressure points will be shown (both the default
pressure check point and the actual pressure measured will appear
alternatively.).
If "FAIL" happens, press the "Auto calibration" key for re-calibration.
After this process is complete, press the "Readout" key for double
check.
Checking the pressures between calibrator and
monitor. If the deviation of First Pressure Points is
under ±3mmHg, the calibrator will automatically
check Second Pressure Points (210/180/150 mmHg)
check point on
Calibrator
check point from Blood
Pressure Monitor
If deviation under ±3mmHg
If deviation over ±3mmHg
Automatic Check 3rd
Pressure Points
Calibrator Screen
Blood Pressure
Monitor Screen
Calibrator Screen
Blood Pressure
Monitor Screen
Press Auto Calibration Mode
(follow instructions on Page 18)
Calibrator Screen
Blood Pressure
Monitor Screen
First
Pressure
Points
If deviation between calibrator and
monitor is under ±3mmHg
Second
Pressure
Points