2. Readout Mode
This feature checks monitor pressure, if you would like to ignore
pressure checking , skip to "Auto Calibration Mode"(page 18) directly.
Press “Readout” to start. “RUN” will appear on LCD and pressure is
checked at 0/295/270/240 mmHg points. If the pressure difference
at 0/295/270/240 mmHg is within ± 3mmHg, the calibrator will rapid
exhaust to 0 mmHg, then go on checking 2nd points. If "FAIL" appears
, the faulty pressure points will be shown (both the default pressure
check point and the actual pressure measured will appear alternatively.).
If "FAIL" appears, press the "Auto calibration" key for re-calibration.
After this process is complete, press the "Readout" key to verify
calibration was successful.