1.  General Information 

 

This User Manual is valid for the Calibration Kit  

 

19CK010-150 (Industrial Version)

 

 
 

1.1 Calibration Kit Normal Use 

 

Calibration  Kits  are  used  as  an  accessory  for  Vector  Network  Analyzers  (VNA)  to  perform 
vector-error correction. Vector-error correction is the process of characterizing systematic errors 
of  the  measurement  system  by  measuring  known  calibration  standards,  and  then 
mathematically  removing  the  effects  of  these  errors  from  subsequent  measurements.  The 
measurement accuracy of the calibrated VNA depends mainly on the quality of the calibration kit 
used. Contributing factors are noise, linearity and temperature stability of the VNA, transmission 
and reflection stability of a test port cable as well as operator experience and care. 

 

VNAs  usually  offer  several  techniques  for  correcting  systematic  measurement  errors.  The 
standards  in  this  calibration  kit  allow  you  to  perform  one-port  OSL  (OSM)  or  two-port  OSLT 
(TOSM) and TRM (TMR) calibrations from DC up to 40 GHz for measurements of components 
with SMP connectors. 

 

RPC-2.92  to  SMP  adapters  are  included  in  this  calibration  kit  to  interface  with  standard  
2.92 mm, K or 3.5 mm test port cables and VNAs.  

 

1.2 Warnings 

 

The  calibration  standards  of  this  calibration  kit  are  mechanically  extremely  sensitive  devices. 
High  accelerations  e.g.  when  falling  down  or  high  forces  to  the  connector fingers  will  damage 
them. 

 

When connecting and disconnecting male and female connectors keep them in a straight line as 
much as possible. Radial misalignment (more than permitted) stresses the fingers of the female 
outer conductor significantly and leads to bent or broken fingers. This results in a very unstable 
or bad RF performance.  
 

 

Connect the calibration standards only to test ports and other devices that have been cleaned 
and  checked  for  gauge,  concentricity  and  integrity  of  the  contact  fingers.  Detailed  procedures 
are described in Chapter 6 and 7. 

 

Tighten the connector coupling nuts of the RPC-2.92 interfaces to the recommended torque with 
a torque  wrench (e.g. Rosenberger 03W021-000) while holding the RPC-2.92 female counter-
part with a wrench (e.g. Rosenberger Combi Wrench 03W021-000). Do not over torque! 

 

Consider  the  power  applied  to  the  Load  calibration  standards.  Maximum  power  limits  for  the 
sum of RF and DC power are defined in the specific data sheets. 

 

It is good lab practice to use at least an ESD wrist strap when handling calibration standards or 
DUTs on a VNA although the calibration standards themselves are not ESD sensitive.  Please 
consult your VNA manual. 

 

Summary of Contents for 19CK010-150

Page 1: ...hnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 16 Dieses Dokument ist urheberrechtlich gesch tzt This document is protec...

Page 2: ...Kit Documentation 1 4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and Mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installat...

Page 3: ...ments of components with SMP connectors RPC 2 92 to SMP adapters are included in this calibration kit to interface with standard 2 92 mm K or 3 5 mm test port cables and VNAs 1 2 Warnings The calibrat...

Page 4: ...plug plug 19S121 S20D3 1 Calibration adaptor jack jack 19K121 K20D3 1 Calibration adaptor plug jack 19S121 K20D3 1 Calibration adaptor plug plug 02S119 S20D3 1 Calibration adaptor jack jack 02K119 K20...

Page 5: ...ing temperature range for which the specifications are valid is defined in the specific data sheets 2 2 Electrical and mechanical Specifications Electrical and mechanical specifications are defined fo...

Page 6: ...lex process Please consult the VNA operating man ual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously loading the...

Page 7: ...ers Sweep type linear logarithmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the dynamic range Swe...

Page 8: ...calibration standards have to be entered manually The Reflect device is derived from the short calibration standard but the L0 L1 L2 and L3 are all set to zero This keeps the symmetry of the Reflect...

Page 9: ...case of re calibration or repair contact your local Rosenberger sales office 5 4 Extent of Rosenberger Re Calibration Services Every calibration kit sent for re calibration or repair the following se...

Page 10: ...that have dents on the mating surfaces will usually also have raised edg es around them and will make less than perfect contact further to this raised edges on mating interfaces will make dents in ot...

Page 11: ...irectly on to the con nector surfaces or use contaminated solvents Use the low pressure compressed air once again to remove any remaining small particles and to dry the surfaces of the connector to co...

Page 12: ...destructive mating and electrical performance of the connector The mechanical gauging of coaxial connectors will detect and prevent the following problems Inner conductor protrusion This may result i...

Page 13: ...6 3 Handling and Storage Install the protective end caps and store the calibration devices in the foam lined storage case when not in use Never store connectors loose in a box or in a desk or bench dr...

Page 14: ...the measurements and even make them unusable Adapting Test Ports The calibration kit comes with two RPC 2 92 to SMP adapters to adapt the test port of the VNA or the 2 92 mm test port cable to SMP in...

Page 15: ...one or more contact la mellae which in turn can result in different measurement results This mated connection is kept free of me chanical stress This mated connection is bent too much and results in...

Page 16: ...mall but the full detent version needs a high force to disconnect Care should be taken not to get hurt during this process and to keep the test port cable and the calibra tion standard under control T...