6.1.3 Cleaning procedure 

 
Items required: 
 

  Low pressure compressed air (solvent free). 

  Cotton swabs. 

  Lint free cleaning cloth. 

  Isopropanol. 

  Illuminated magnifier or microscope (factor 5 to 10) 

 
 

NOTE:  Isopropanol that contains additives should not be used for cleaning connectors 

as it may cause damage to plastic dielectric support beads in coaxial and micro-
wave  connectors.  It  is  important  to  take  all  necessary  safety  precautions  when 
using chemicals or solvents. 

 

  Remove  loose  particles  on  the  mating  surfaces  and  threads  etc.  using  low-pressure  com-

pressed air. 

  Clean surfaces using Isopropanol on cotton swabs or lint free cloth. Use only sufficient sol-

vent to clean the surface. When using swabs or lint free cloth,  use the least possible pres-
sure  to  avoid  damaging  connector  surfaces.  Do  not  spray  solvents  directly  on  to  the  con-
nector surfaces or use contaminated solvents. 

  Use  the  low-pressure  compressed  air  once  again  to  remove  any  remaining  small  particles 

and to dry the surfaces of the connector to complete the cleaning process before using the 
connector. 

 

6.1.4 Cleaning connectors on static sensitive devices 

 
Special  care  is  required  when  cleaning  connectors  on  test  equipment  containing  electrostatic 
sensitive devices. When cleaning such connectors always wear at least a grounded wrist strap 
and observe correct procedures.  

 

6.2 Interface Dimensions and Gauging 

 
It is of the utmost importance that connectors do not damage the test equipment interfaces to 
which  they  are  connected.  Poor  performance  of  many  coaxial  devices  and  cable  assemblies 
can  often  be  traced  to  poor  construction  and  non-compliance  with  the  mechanical  specifica-
tions. The mechanical gauging of connectors is essential to ensure correct fit and to achieve the 
best  performance.  This  means  that  all  coaxial  connectors  fitted  on  all  equipment,  cables  and 
terminations  etc.  should  be  gauged  on  a  regular  basis  in  order  to  detect  any  out  of  tolerance 
conditions that may impair the electrical performance. 
 

6.2.1 Gauging connectors 

 
A  connector should  be gauged  before  it  is  used  for the very  first  time  or  if someone  else  has 
used the device to which it is connected. 

Summary of Contents for 19CK010-150

Page 1: ...hnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 16 Dieses Dokument ist urheberrechtlich gesch tzt This document is protec...

Page 2: ...Kit Documentation 1 4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and Mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installat...

Page 3: ...ments of components with SMP connectors RPC 2 92 to SMP adapters are included in this calibration kit to interface with standard 2 92 mm K or 3 5 mm test port cables and VNAs 1 2 Warnings The calibrat...

Page 4: ...plug plug 19S121 S20D3 1 Calibration adaptor jack jack 19K121 K20D3 1 Calibration adaptor plug jack 19S121 K20D3 1 Calibration adaptor plug plug 02S119 S20D3 1 Calibration adaptor jack jack 02K119 K20...

Page 5: ...ing temperature range for which the specifications are valid is defined in the specific data sheets 2 2 Electrical and mechanical Specifications Electrical and mechanical specifications are defined fo...

Page 6: ...lex process Please consult the VNA operating man ual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously loading the...

Page 7: ...ers Sweep type linear logarithmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the dynamic range Swe...

Page 8: ...calibration standards have to be entered manually The Reflect device is derived from the short calibration standard but the L0 L1 L2 and L3 are all set to zero This keeps the symmetry of the Reflect...

Page 9: ...case of re calibration or repair contact your local Rosenberger sales office 5 4 Extent of Rosenberger Re Calibration Services Every calibration kit sent for re calibration or repair the following se...

Page 10: ...that have dents on the mating surfaces will usually also have raised edg es around them and will make less than perfect contact further to this raised edges on mating interfaces will make dents in ot...

Page 11: ...irectly on to the con nector surfaces or use contaminated solvents Use the low pressure compressed air once again to remove any remaining small particles and to dry the surfaces of the connector to co...

Page 12: ...destructive mating and electrical performance of the connector The mechanical gauging of coaxial connectors will detect and prevent the following problems Inner conductor protrusion This may result i...

Page 13: ...6 3 Handling and Storage Install the protective end caps and store the calibration devices in the foam lined storage case when not in use Never store connectors loose in a box or in a desk or bench dr...

Page 14: ...the measurements and even make them unusable Adapting Test Ports The calibration kit comes with two RPC 2 92 to SMP adapters to adapt the test port of the VNA or the 2 92 mm test port cable to SMP in...

Page 15: ...one or more contact la mellae which in turn can result in different measurement results This mated connection is kept free of me chanical stress This mated connection is bent too much and results in...

Page 16: ...mall but the full detent version needs a high force to disconnect Care should be taken not to get hurt during this process and to keep the test port cable and the calibra tion standard under control T...

Reviews: