7. Connections 

 
Good connections require a skilled operator. The most common cause of measurement error is 
bad connections. The following procedures illustrate how to make good connections. 
The wearing of protective gloves is strongly recommended for every single connection. Even the 
smallest amount of skin lipids/particles that gets into the connector interface can have a strong 
influence on the measurements and even make them unusable. 
 
 

Adapting Test Ports 

 

The calibration kit comes with two RPC-2.92 to SMP adapters to adapt the test port of the VNA 
or the 2.92 mm test port cable to SMP interfaces. 
 

How to connect  

 

Pick up the calibration standard. Make sure that the test port cable and the calibration standard are 
aligned with each other. Carefully guide it to the test port cable until you are fully centred and you 
feel the male and female SMP connectors touching each other. Now the force can be gradually in-
creased until the SMP connection snaps into place. 

 

 

 

 

The test port cable and the calibration standard 
are aligned with each other.

 

 

The test port cable and the calibration standard 
are not aligned with each other. A strong me-
chanical stress will be caused.

 

 

 

 

Draft 

Date 

Approved 

Date 

  Rev. 

Engineering change number

 

Name 

Date 

Herbert Babinger 

24/04/12 

Markus Hantschel 

21/12/20   

a00 

20-1058 

Lars Ramtke 

21/12/20 

Summary of Contents for 19CK010-150

Page 1: ...hnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 16 Dieses Dokument ist urheberrechtlich gesch tzt This document is protec...

Page 2: ...Kit Documentation 1 4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and Mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installat...

Page 3: ...ments of components with SMP connectors RPC 2 92 to SMP adapters are included in this calibration kit to interface with standard 2 92 mm K or 3 5 mm test port cables and VNAs 1 2 Warnings The calibrat...

Page 4: ...plug plug 19S121 S20D3 1 Calibration adaptor jack jack 19K121 K20D3 1 Calibration adaptor plug jack 19S121 K20D3 1 Calibration adaptor plug plug 02S119 S20D3 1 Calibration adaptor jack jack 02K119 K20...

Page 5: ...ing temperature range for which the specifications are valid is defined in the specific data sheets 2 2 Electrical and mechanical Specifications Electrical and mechanical specifications are defined fo...

Page 6: ...lex process Please consult the VNA operating man ual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously loading the...

Page 7: ...ers Sweep type linear logarithmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the dynamic range Swe...

Page 8: ...calibration standards have to be entered manually The Reflect device is derived from the short calibration standard but the L0 L1 L2 and L3 are all set to zero This keeps the symmetry of the Reflect...

Page 9: ...case of re calibration or repair contact your local Rosenberger sales office 5 4 Extent of Rosenberger Re Calibration Services Every calibration kit sent for re calibration or repair the following se...

Page 10: ...that have dents on the mating surfaces will usually also have raised edg es around them and will make less than perfect contact further to this raised edges on mating interfaces will make dents in ot...

Page 11: ...irectly on to the con nector surfaces or use contaminated solvents Use the low pressure compressed air once again to remove any remaining small particles and to dry the surfaces of the connector to co...

Page 12: ...destructive mating and electrical performance of the connector The mechanical gauging of coaxial connectors will detect and prevent the following problems Inner conductor protrusion This may result i...

Page 13: ...6 3 Handling and Storage Install the protective end caps and store the calibration devices in the foam lined storage case when not in use Never store connectors loose in a box or in a desk or bench dr...

Page 14: ...the measurements and even make them unusable Adapting Test Ports The calibration kit comes with two RPC 2 92 to SMP adapters to adapt the test port of the VNA or the 2 92 mm test port cable to SMP in...

Page 15: ...one or more contact la mellae which in turn can result in different measurement results This mated connection is kept free of me chanical stress This mated connection is bent too much and results in...

Page 16: ...mall but the full detent version needs a high force to disconnect Care should be taken not to get hurt during this process and to keep the test port cable and the calibra tion standard under control T...