Figure 5-6. Fault 6, High Case Temp
f.
Fault 6, High Case Temp. The HEATER LED
flashes three times, pauses for three seconds, and
repeats (Figure 5-6).
1. If the case temperature exceeds 185°F
(85°C), the temperature control will shut off
and the 4-20 mA signal output will go to the
default value.
2. This signifies that the environment where
the Oxymitter 4000 is installed exceeds the
ambient temperature requirements or that
heat due to convection is causing case tem-
perature to rise above the limit.
3. Placing a spool piece between the stack
flange and the Oxymitter 4000 flange may
eliminate this problem.
4. If a spool piece does not solve the problem,
relocation is the only solution.