Figure 5-3. Fault 3, Reversed Thermocouple
c.
Fault 3, Reversed Thermocouple. The
HEATER T/C LED flashes three times, pauses
for three seconds, and repeats (Figure 5-3).
1. Using a multimeter, measure TP3+ to TP4-.
2. If the reading is negative, the thermocouple
wiring is reversed.
3. Check red and yellow wires in the J1
connector for the proper placement.
4. If the wiring is correct, the fault is in the PC
board. See paragraph 6-5b, Electronic As-
sembly Replacement.