5.1.2 P
HYSICAL
C
ONFIGURATION OF
D
ETECTOR
/M
AGNET
A
SSEMBLY
As shown in the exploded view of Figure 5-3A, the detector/magnet assembly consists
of three major subassemblies: the magnet assembly, the detector assembly, and the
optical bench assembly.
The magnet assembly includes a sample preheating coil. It is connected into the
sample line, upstream from the detector, and is heated to approximately the same
temperature as the detector assembly.
F
IGURE
5-3. D
ETECTOR
/M
AGNET
A
SSEMBLY
Detector Assembly
Mounting Screws (2)
Optical Bench Assembly
Magnet Assembly
Sample Outlet Tube
Sample Inlet Tube
Sample Pre-Heating Coil
Connector J12
Dual Photocell
Connector J12
Lamp Retaining
Set Screw
Photocell
Lock Screws (2)
Source Lamp
Assembly
Lamp Viewing Hole
➊
Dual Photocell
➋
Integral Temperature
Sensor (RT1)
➋
Integral 5-Micron
Diffusion Screen
➋
Test Body
➋
Integral Heater (HR2)
Mirror
➊
Optical Bench Assembly
➋
Detector Assembly
➊
Source Lamp
A. Exploded View of Detector/Magnet Assembly
B. Sectional Top View of Optical Bench
and Detector Assemblies
C. Exploded View of Optical Bench
Assembly
Summary of Contents for 755
Page 1: ...Rosemount Analytical MODEL 755 OXYGEN ANALYZER INSTRUCTION MANUAL 748183 H ...
Page 16: ...MODEL 755 OXYGEN ANALYZER Rosemount Analytical 748183 H November 1997 P 8 NOTES ...
Page 22: ...MODEL 755 OXYGEN ANALYZER 1 6 748183 H Rosemount Analytical November 1997 NOTES ...
Page 40: ...MODEL 755 OXYGEN ANALYZER 2 18 748183 H Rosemount Analytical November 1997 NOTES ...
Page 50: ...MODEL 755 OXYGEN ANALYZER 3 10 748183 H Rosemount Analytical November 1997 NOTES ...
Page 62: ...MODEL 755 OXYGEN ANALYZER 5 10 748183 H Rosemount Analytical December 1997 NOTES ...
Page 90: ... blank ...
Page 92: ... blank ...
Page 94: ... blank ...