performed on the range of highest sensitivity, i.e., most narrow span. Preferably
upscale calibration should be performed on the range that will be used during sample
analysis. In some applications however, it may be desirable to perform upscale
calibration on a range of higher sensitivity, i.e., more narrow span, and then move the
% RANGE Switch to the desired operating range. For example, if the operating range
is to be 0 to 50% oxygen upscale calibration may be performed on the 0 to 25% range,
to permit use of air as the upscale standard gas.
It is necessary to calibrate the instrument at the same pressure that will be used during
subsequent operation and to maintain this pressure during operation.
The preferred calibration method uses both a downscale and an upscale standard gas,
as described in Section 3.2.1. An alternative method using an upscale standard gas
only is described in Section 3.2.2.
F
IGURE
3-1. C
ONTROL
B
OARD
- A
DJUSTMENT
L
OCATIONS
Summary of Contents for 755
Page 1: ...Rosemount Analytical MODEL 755 OXYGEN ANALYZER INSTRUCTION MANUAL 748183 H ...
Page 16: ...MODEL 755 OXYGEN ANALYZER Rosemount Analytical 748183 H November 1997 P 8 NOTES ...
Page 22: ...MODEL 755 OXYGEN ANALYZER 1 6 748183 H Rosemount Analytical November 1997 NOTES ...
Page 40: ...MODEL 755 OXYGEN ANALYZER 2 18 748183 H Rosemount Analytical November 1997 NOTES ...
Page 50: ...MODEL 755 OXYGEN ANALYZER 3 10 748183 H Rosemount Analytical November 1997 NOTES ...
Page 62: ...MODEL 755 OXYGEN ANALYZER 5 10 748183 H Rosemount Analytical December 1997 NOTES ...
Page 90: ... blank ...
Page 92: ... blank ...
Page 94: ... blank ...