748183-H
7-3
Rosemount Analytical
December 1997
V
OLTAGE
T
EST
M
EASUREMENTS
B TO A
C OR D TO
GROUND
DIAGNOSIS
CORRECTIVE ACTION
-
+
Normal
NA
+
+
U1 or U2 defective
Replace case board
-
-
U1 or U2 defective
Replace case board
+
-
Detector defective
Check detector per Section 7.3
If polarities are correct, set front-panel SPAN potentiometer R4 at maximum. The
output at pin 6 of U3 should be 7.5 VDC. Pins 2 and 3 of U4 should both be at 0.465
VDC, resulting in 5 VDC at the output of Q1 .
Checkout of the case circuit board is now complete.